Computational methodology for the evaluation of tangential stresses and torsional moments in heterogeneous total hip arthroplasty assemblies
M Mejía-Rodríguez, D Jiménez, O A González-Estrada, J D Sierra-Villegas, A Velásquez
Key points
From the paper's conclusion- The numerical results revealed peak von Mises stresses in the liner between 5.50 MPa and 5.87 MPa, locking tab micromotions up to 6.56 µm, and resultant torsional moments varying from −24.79 N·mm to −162.06 N·mm.
Abstract
The dissociation of the polyethylene liner within acetabular cups is a rare but catastrophic event involving a sudden loss of joint congruity in total hip arthroplasty. While traditionally attributed to mechanical impingement or progressive wear of the locking mechanism, this study shows that polyethylene rotation and the mechanical demand on the anchoring system are strongly influenced by tangential forces at the head-liner interface. A finite element methodology coupled with a mathematical vectorial model was executed on digitized commercial components subjected to a 3 k factorial experimental design under a 3139 N physiological joint contact load. The numerical results revealed peak von Mises stresses in the liner between 5.50 MPa and 5.87 MPa, locking tab micromotions up to 6.56 µm, and resultant torsional moments varying from −24.79 N·mm to −162.06 N·mm. Heterogeneous cross-pairings exhibited substantial variations in interfacial contact pressure and rotational demand, confirming that rotational instability can be triggered independently of mechanical neck-rim impingement. This computational approach provides an objective tool to assess mechanical viability of mixed prosthetic assemblies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Early Catastrophic Failure of Highly-Cross-Linked Polyethylene Liner Due to Acetabular Shell Screw Placement: A Case Report.Cureus · 2026
- Catastrophic acetabular wear following polyethylene liner dissociation with locking-tab damage after total hip arthroplasty: a case report.Frontiers in surgery · 2026
- Peg Shearing as a Method of Failure in Highly Cross-Linked Polyethylene Patellar Components Used in Patellofemoral Arthroplasty.Arthroplasty today · 2026
- Navigation-Assisted and Robotic Total Hip, Knee, and Unicompartmental Arthroplasty: Reconciling Positioning Accuracy with Biomaterial Performance and Long-Term Implant Survival.Bioengineering (Basel, Switzerland) · 2026
- Newer Generation DePuy INHANCE™ Modular Anatomic Convertible Polyethylene Glenoid and Platform Baseplate have Improved Resistance to Micromotion and Dissociation at the Polyethylene-Metal Junction.Journal of shoulder and elbow surgery · 2026
- Asymmetric Metaphyseal Cones for AORI Type 2 Medial Defects in Tibial Plateau Fractures during Acute TKAJournal of Orthopaedic Research® · 2026