Seamless vital sign monitoring with fiber optic sensing technology: validating the sitting SMART MAT for community health in static and mobile conditions.
Yan Yi Sim, Darryl J Q Chan, Yong Xiang Phang, Marc S L Liem, Meredith T Yeung
Health and Social Sciences Cluster, Singapore Institute of Technology, Singapore, Singapore.
Key points
From the paper's conclusion- The current SMART MAT is a viable, integrated multimodal platform. HR, RR, and DBP are comparable to clinical reference standards, while SBP requires careful interpretation. By accommodating a functional sitting posture, the contactless system reduces immobility-related complications and demonstrates high feasibility for continuous, real-world monitoring.
Abstract
BACKGROUND: Continuous vital sign monitoring is critical but often labor-intensive. Fiberoptic technologies, like the SMART MAT, offer non-invasive assessment. Adapting this for a seated posture addresses clinical needs by promoting safe patient mobilization and reducing complications associated with prolonged bed rest. OBJECTIVES: To validate the accuracy of the sitting SMART MAT for respiratory rate (RR), heart rate (HR), oxygen saturation (SpO2), and systolic and diastolic blood pressure (SBP/DBP) against established clinical reference standards. METHODS: This cross-sectional study compared sitting SMART MAT readings with clinical reference measurements under stationary and mobile wheelchair conditions. Accuracy was assessed using Mean Absolute Error (MAE), Mean Absolute Percentage Error (MAPE), Bland-Altman Limits of Agreement (LoA), repeated-measures ANOVA, and effect size. RESULTS: Data from 130 participants were analyzed. HR readings demonstrated high accuracy (MAE = 3.67-4.81 bpm) with tight LoA. RR error was minimal (MAE = 2.85-2.86 breaths/min) statically but increased during motion (MAE = 5.03-5.21 breaths/min). SpO2 remained highly stable (MAE = 1.08-1.21%), well within standard clinical tolerance. SBP showed solid central accuracy (MAE = 6.48-7.51 mmHg) but wider LoA occasionally exceeded the ±5 mmHg threshold, functioning best as a continuous trend monitor. DBP variance was negligible (MAE = 4.65-5.46 mmHg). CONCLUSION: The current SMART MAT is a viable, integrated multimodal platform. HR, RR, and DBP are comparable to clinical reference standards, while SBP requires careful interpretation. By accommodating a functional sitting posture, the contactless system reduces immobility-related complications and demonstrates high feasibility for continuous, real-world monitoring.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy of cuffless blood pressure devices compared with invasive arterial line monitoring in high-acuity settings: a random-effects meta-analysisEuropean Heart Journal · 2026
- Reduction of Motion Artifacts in Cardiorespiratory Vital Signs as Oxygen Saturation and Non-Invasive Blood Pressure Through Redundant Denoising and Adaptive Filtering Methods for Wearable Healthcare Monitoring SystemsCardiovascular Engineering and Technology · 2026
- Pulse Oximetry Versus Arterial Co-Oximetry During Exercise: A Systematic Review of Agreement and Sensor Technology Reporting Gaps.Biosensors · 2026
- Performance validation of a new finger-root-adapted ring-shaped oximeter sensor during stepwise steady-state inspired hypoxia.Journal of clinical monitoring and computing · 2026
- Comparing noninvasive blood pressure monitoring technologies: the challenge of between-arm differences.British journal of anaesthesia · 2026
- A Wearable Morphic Sensor for Unobtrusive Heart Rate and Respiratory Rate Monitoring in ICU Patients: A Pilot Validation Study.Sensors (Basel, Switzerland) · 2026