Reduction of Motion Artifacts in Cardiorespiratory Vital Signs as Oxygen Saturation and Non-Invasive Blood Pressure Through Redundant Denoising and Adaptive Filtering Methods for Wearable Healthcare Monitoring Systems
Fabián Andrés Castaño, Alher Mauricio Hernández
Universidad de Antioquia
Key points
From the paper's conclusion- The results demonstrate significant improvements in signal quality and physiological parameter estimation compared with conventional denoising approaches, supporting the potential use of the method in ambulatory cardiovascular monitoring applications.
Abstract
Abstract Background and objective Chronic cardiovascular diseases have motivated the development of wearable systems capable of continuously monitoring physiological variables such as oxygen saturation (SpO $$_2$$ 2 ) and non-invasive blood pressure (NIBP). However, motion artifacts remain one of the main limitations affecting the reliability of these measurements in ambulatory environments. This article presents a novel framework for motion artifact reduction in photoplethysmography (PPG) and NIBP signals based on redundant sensing, inertial measurements, ECG-guided temporal synchrony, and multichannel signal processing. Methods The proposed framework combines independent component analysis and recursive least-squares adaptive filtering to separate physiological information from motion-induced interference. Performance was compared with conventional finite impulse response (FIR) filtering and wavelet shrinkage (WS) methods using signal-to-noise ratio (SNR), weighted distortion assessment (WDA), oxygen saturation estimation, and blood pressure estimation metrics. Results The proposed method achieved statistically significant improvements in signal quality and physiological parameter estimation. Average SNR improved from $$-5.93\pm 5.05$$ - 5.93 ± 5.05 dB to $$-3.43\pm 2.02$$ - 3.43 ± 2.02 dB in the red PPG channel ( $$p=0.02$$ p = 0.02 ), while the WDA index increased from 0.48 to 0.73. Oxygen saturation estimation improved from $$80.87\pm 9.09\%$$ 80.87 ± 9.09 % in motion-contaminated signals to $$94.16\pm 2.72\%$$ 94.16 ± 2.72 % after denoising ( $$p=0.04$$ p = 0.04 ), approaching resting measurements ( $$95.23\pm 2.31\%$$ 95.23 ± 2.31 % ). For NIBP signals, statistically significant improvements were also observed ( $$p=0.045$$ p = 0.045 ), particularly for systolic pressure estimation, although diastolic pressure remained affected by residual motion-related distortions. Conclusions The proposed framework improves the robustness of wearable PPG and NIBP monitoring under motion conditions through the combined use of sensor redundancy, ECG-guided temporal coupling, inertial measurements, and adaptive multichannel signal processing. The results demonstrate significant improvements in signal quality and physiological parameter estimation compared with conventional denoising approaches, supporting the potential use of the method in ambulatory cardiovascular monitoring applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Signal quality-guided adaptive denoising for robust remote photoplethysmography.Biomedical engineering letters · 2026
- A Novel Cluster-Based Approach for PPG Signal Preprocessing: Motion Artifact Filtering and Ambient Light Correction.Sensors (Basel, Switzerland) · 2026
- A Context-Aware Framework for Optimal Filter Parameter Selection: Enhancing PRV Estimation in Wearable Wrist PPG SignalsProceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies · 2026
- An Agent Architecture with Tools and Memory for Selective Correction of Wrist PPG Heart Rate EstimatesarXiv · 2026
- Methods for Optimizing Photoplethysmography Signal Quality.Sensors (Basel, Switzerland) · 2026
- Robust synchronization of cardiac-linked biosensors under peak detection failures: a modality-agnostic approach via structural repair and Huber-IRLS regressionMeasurement Science and Technology · 2026