Pulse width-dependent Monte Carlo source modelling for ultra-high dose rate electron beams.
Angela Maria Henao Isaza, Ivan Lopez Paz, Consuelo Guardiola, Johan Sebastián Moreno, Sophie Heinrich
Instituto de Microelectronica de Barcelona, Carrer dels Til·lers, Campus de la UAB · Bellaterra, Cerdanyola del Vallès, 08193, Spain. · Instituto de Microelectronica de Barcelona, Carrer dels Til·lers s/n, Campus de la UAB · Bellaterra, Cerdanyola del Vallès, 08193, Spain.
Key points
From the paper's conclusion- A GATE 10 MC-based source model of the ElectronFLASH LINAC was developed for both UHDR (PW-dependent) and conventional modes. Despite PW-dependent energy variations, no significant dosimetric differences were observed between PWs in FLASH in the preclinical mouse model. The model provides a reliable tool for optimization of preclinical irradiation setups for FLASH biological studies in the absence of a treatment planning system (TPS).
Abstract
OBJECTIVE: The emergence of ultra-high dose rate (UHDR) electron beams has highlighted the need for accurate Monte Carlo (MC) source models after recent measurements revealed pulse width-dependent discrepancies in dose profiles. This work aims to develop and validate a MC source model for conventional and FLASH modes using GATE 10, including pulse width (PW) dependence. APPROACH: Percentage depth dose (PDD) curves and lateral dose profiles were measured in water using a flashDiamond detector for PWs of 1 -5 µs. These data were used to optimized the parametric source model through χ 2 minimization and gamma analysis. Validation was performed using a mouse collimator and PMMA slabs. Relative dose distributions and cumulative dose-volume histograms (cDVHs) were computed in a CT-based voxelised mouse, with and without bolus. MAIN RESULTS: The optimized FLASH source model features a dual-peak energy spectrum with a PW-dependent energy component (7.5-7.8 MeV for 1-5 µs) and low-energy component probability. In conventional mode, a two-component spectrum (1.5 ± 1.0 and 6.8 ± 1.5 MeV) was used. Simulations showed good agreement with measurements (gamma-index 2 mm/2% and < 5% differences for R 50 , R 90 and R p ). Differences in the mouse lung volume receiving at least 95% of the maximum dose (V95 lung ) were 1% across FLASH PWs, increasing to 5% and 12 % in CONV without and with bolus, respectively. Bolus increased V95 lung in FLASH by 17%, indicating improved target coverage. SIGNIFICANCE: A GATE 10 MC-based source model of the ElectronFLASH LINAC was developed for both UHDR (PW-dependent) and conventional modes. Despite PW-dependent energy variations, no significant dosimetric differences were observed between PWs in FLASH in the preclinical mouse model. The model provides a reliable tool for optimization of preclinical irradiation setups for FLASH biological studies in the absence of a treatment planning system (TPS).
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pulse-width-specific phase space modeling and reference beam characterization for UHDR electron LINAC in FLASH-RTPhysics in Medicine and Biology · 2026
- VHEE beam parameter extraction from depth-dose and lateral profile measurements: a Monte Carlo parametric study.Radiological physics and technology · 2026
- Toward dynamic Monte Carlo source modeling: measurement variability and primary electron source-width sensitivity in very small photon fields.Frontiers in oncology · 2026
- Ultra-high-dose-rate pulsed irradiation and the conditional FLASH effect: dosimetry, multiscale biology, and clinical translation.Frontiers in oncology · 2026
- Breathing-phase selection for gated proton therapy in lung cancer towards temporally targeted ultra-high dose rate delivery.Physics and imaging in radiation oncology · 2026
- Monte-Carlo simulations of Cherenkov Fluence for FLASH radiotherapy dosimetryJournal of Physics: Conference Series · 2026