Monte Carlo evaluation of small-field output correction factors of a microDiamond detector in flattened and flattening-filter-free photon beams for stereotactic radiotherapy.
Hamza Sekkat, Oussama El Mouden, Abdellah Khallouqi, Youssef Madkouri, Abdellah Halimi, Omar El Rhazouani
Laboratory of Sciences and Engineering of Biomedicals, Biophysics and Health, Higher Institute of Health Sciences, Hassan 1 st University,, Settat, Marocco. · Laboratory of Electronic Systems, Information Processing, Mechanics and Energetics, Faculty of Sciences, Ibn Tofail University, Kenitra, Morocco. youssef.madkouri@uit.ac.ma.
Key points
From the paper's conclusion- This study derived machine-specific IAEA TRS-483 output correction factors for a PTW 60,019 microDiamond detector on an Elekta Versa HD linac for 6 MV and 10 MV photons in both flattened filter (FF) and FFF modes, using square fields down to 1 × 1 cm².
Abstract
Small-field dosimetry is clinically challenging because of volume averaging, loss of lateral charged-particle equilibrium, and beam-quality changes, especially for multi-leaf collimator (MLC)-defined fields and flattening-filter-free (FFF) beams. This study derived machine-specific IAEA TRS-483 output correction factors for a PTW 60,019 microDiamond detector on an Elekta Versa HD linac for 6 MV and 10 MV photons in both flattened filter (FF) and FFF modes, using square fields down to 1 × 1 cm². A detailed GATE/Geant4 Monte Carlo model of the treatment head (including the Agility MLC) was validated against 10 × 10 cm² percentage depth doses (PDDs) (mean absolute differences 0.20-0.29% points) and 1 × 1 cm² profiles (FWHM differences < 0.06 mm). Uncorrected MC-measurement output factor deviations at 1 × 1 cm² reached + 1.03% (6 MV), + 1.15% (10 MV), + 1.66% (6 MV FFF), and + 2.08% (10 MV FFF), but stayed ≤ 0.3% for fields ≥ 3 × 3 cm². Corresponding output correction factors at 1 × 1 cm² were 1.0103, 1.0115, 1.0166, and 1.0208, approaching unity (± 0.2%) for ≥ 5 × 5 cm². Agreement with TRS-483 reference tables was within 0.6% at 1 × 1 cm² and < 0.4% at 2 × 2 cm². Film and MicroSilicon checks agreed within 1-2% at 2 × 2 cm², but differed by 13-15% at 1 × 1 cm² for 10 MV, reinforcing the need for machine-specific factors, particularly for 1 × 1 cm² and FFF beams. These findings indicate that small-field detector correction should be regarded as specific to the treatment-unit, beam mode, and field configuration rather than as a universally transferable quantity. In clinical implementation, Monte Carlo-based correction should therefore be integrated with independent verification when dosimetry approaches the limits of detector and field-size resolution.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of percentage depth dose measurements obtained with three detector technologies and Monaco TPS calculations in a 1 × 1 cm² small photon field.Biomedical physics & engineering express · 2026
- Characterization and Validation of MicroDiamond Detectors in Patient-specific Quality Assurance of Magnetic Resonance-guided Stereotactic Radiotherapy.Journal of medical physics · 2026
- Performance evaluation of a new type of commercial silicon detectors for photon beam dosimetryJournal of Applied Clinical Medical Physics · 2026
- Dosimetric comparison of 6-MV flattened and flattening filter-free photon beams using Monte Carlo simulationNuclear Engineering and Technology · 2026
- Toward dynamic Monte Carlo source modeling: measurement variability and primary electron source-width sensitivity in very small photon fields.Frontiers in oncology · 2026
- Comparative dosimetric analysis of flattening-filter and flattening-filter-free photon beams in radiotherapy based on TRS-398 calibration protocol.Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation Oncology · 2026