Gantry-period-guided modulation factor selection in helical tomotherapy: effects on delivery efficiency, dosimetry and verification accuracy.
Xiaoli Jin, Tianhui Gao, Xiangqian Liu, Hailiang Zhang, Xin Yang, Xia Zhang
Department of Oncology, Shanxi Bethune Hospital, Shanxi Academy of Medical Sciences, Third Hospital of Shanxi Medical University, Tongji Shanxi Hospital, Taiyuan, China. · Department of Biomedical Engineering, Shanxi Medical University, Taiyuan, China.
Key points
From the paper's conclusion- The MF immediately before gantry-period plateau entry may provide a practical reference point for limiting MF reduction. In this pilot cohort, this gantry-period-guided approach helped maintain plan quality and avoid reductions in modulation that offered little efficiency benefit. Further studies across additional planning configurations and disease sites will be needed to confirm its broader applicability.
Abstract
PURPOSE: To explore a gantry-period-guided strategy for identifying a lower useful modulation factor (MF) in helical tomotherapy under fixed field width and pitch. METHODS: Twenty previously treated cases from head, head-and-neck, abdomen and pelvis were replanned with a field width of 2.5 cm and a pitch of 0.287. MF was initialized at 5.0 and reduced by 0.2 with unchanged objectives. At each level, plans were reoptimized and clinically reviewed. MFlimit was defined as the last clinically acceptable MF before the step-down sequence entered the mechanical gantry-period plateau (~11.8 s); MFmin was the lowest clinically acceptable MF identified after further reduction. Gantry period, delivery time, leaf-open-time (LOT) metrics, dosimetric indices and pretreatment gamma pass rates were compared. RESULTS: All cases entered the efficiency-limited pathway. MFlimit ranged from 1.9 to 2.8. MF reduction produced a biphasic gantry-period response, with an initial decrease followed by a mechanical plateau. Further reduction from MFlimit to MFmin did not meaningfully reduce delivery time. LOTmax decreased and LOTmean remained relatively stable, whereas Tgap, defined as the temporal gap between the available projection time and LOTmax, increased markedly at MFmin, indicating unused projection time after plateau entry. Compared with MFmin, MFlimit was associated with improved PTV homogeneity and conformity, more favorable exploratory pooled OAR dose trends, and higher 1%/1 mm gamma pass rates, while 3%/2 mm and 2%/2 mm pass rates were similar. CONCLUSIONS: The MF immediately before gantry-period plateau entry may provide a practical reference point for limiting MF reduction. In this pilot cohort, this gantry-period-guided approach helped maintain plan quality and avoid reductions in modulation that offered little efficiency benefit. Further studies across additional planning configurations and disease sites will be needed to confirm its broader applicability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Association of planned-recorded sinogram difference metrics with treatment planning parameters and gamma pass rates in helical tomotherapy.Radiological physics and technology · 2026
- Log file-based treatment efficiency analysis of TrueBeam and Halcyon delivery systems.Journal of applied clinical medical physics · 2026
- Development and validation of a DICOM-based prediction model for instantaneous dose rate and delivery time in VMAT.Radiological physics and technology · 2026
- Dosimetric and Treatment Delivery Efficiency Comparison of Two Ring-gantry Linear Accelerator Platforms for Liver Stereotactic Body Radiotherapy.Journal of medical physics · 2026
- TrueBeam vs Halcyon: Optimizing dual-LINAC scheduling through comparative dosimetric analysis.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- Megavoltage photon minibeam production through the embedded multi-leaf collimator of a clinical linac: A dosimetric characterization.Medical physics · 2026