Analytical validation of capillary blood Krebs von den Lungen-6 measurement using a high- sensitivity chemiluminescent immunoassay in pediatric patients.
Chengyuan Zhao, Lingyun Shi, Dan Cao, Xianwei Zhang, Tong Cao
Department of Clinical Laboratory Medicine, Children's Hospital of Nanjing Medical University, Nanjing, China. · Nanjing Poclight Biotechnology Company Limited, Nanjing, China.
Key points
From the paper's conclusion- High-sensitivity CLIA on the C5000 analyzer supports the analytical feasibility of KL-6 measurement in pediatric capillary blood. HCT correction improved agreement with venous plasma measurements, although further validation is required before the two sample types can be considered interchangeable for individual clinical assessment and before formal reference intervals can be established.
Abstract
OBJECTIVE: This study evaluates the analytical validity of capillary blood for Krebs von den Lungen-6 (KL-6) detection using a high-sensitivity chemiluminescent immunoassay (CLIA) on the C5000 analyser and assesses agreement between capillary plasma and venous plasma measurements, with emphasis on methodological feasibility rather than clinical diagnostic performance. METHODS: Analytical performance was assessed in terms of repeatability, accuracy and method comparison. Precision was evaluated using pediatric venous plasma samples at three concentration levels, each tested in 20 replicates. Accuracy was verified using manufacturer standards and external quality controls. Comparative analysis involved 80 paired venous plasma and capillary blood samples, with hematocrit (HCT)-based correction applied to address dilutional effects. Agreement between venous and capillary plasma measurements was evaluated using Passing-Bablok regression, Pearson correlation and Bland-Altman analysis. Pediatric and adult KL-6 levels were also profiled to evaluate age-specific differences. RESULTS: The C5000 analyzer demonstrated excellent repeatability, with coefficients of variation below 3% across all concentration levels. Accuracy assessment showed minimal bias relative to the standards and controls. Passing-Bablok regression indicated proportional bias before HCT correction, with a slope of 0.769 (95% CI, 0.550-0.945). After correction, the slope was 1.004 (95% CI, 0.882-1.162) and the intercept was -7.03 U/mL (95% CI, -26.97 to 9.76). The mean capillary-minus-venous bias decreased from -48.32 to -3.47 U/mL, and the Pearson correlation coefficient increased from 0.958 to 0.966. Pediatric KL-6 levels were significantly lower than adult levels (p < 0.001), underscoring the need for pediatric-specific exploratory reference estimates and future diagnostic validation. CONCLUSION: High-sensitivity CLIA on the C5000 analyzer supports the analytical feasibility of KL-6 measurement in pediatric capillary blood. HCT correction improved agreement with venous plasma measurements, although further validation is required before the two sample types can be considered interchangeable for individual clinical assessment and before formal reference intervals can be established.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Capillary Blood Collection Methods: An assessment of different collections technologies compared to traditional venous sampling.Annals of clinical biochemistry · 2026
- Evaluation of point-of-care complete blood count and C-reactive protein testing for routine use in a pediatric emergency department.European journal of pediatrics · 2026
- Extended in-use stability of Krebs von den Lungen-6 (KL-6) reagents, quality controls, and calibrators under refrigerated conditionsInternational Medical Research Journal · 2026
- B-295 Comparison of capillary blood to venous blood for use with the Vitamin D Total II assay on the Atellica IM AnalyzerClinical Chemistry · 2026
- Evaluation of an Upper Arm Capillary Blood Collection Device Compared with Venous and Fingerstick Sampling.The journal of applied laboratory medicine · 2026
- To Convert or Not to Convert? Official International Association of Therapeutic Drug Monitoring and Clinical Toxicology Guideline: Considerations and Recommendations for Converting Capillary Blood Microsampling Concentrations to Plasma ConcentrationsTherapeutic Drug Monitoring · 2025