B-295 Comparison of capillary blood to venous blood for use with the Vitamin D Total II assay on the Atellica IM Analyzer

Agostino Palladino, Brandon Czap, Nancy Carmosino, Christopher DiPasquale, Eric Olson, Roy Barr, Hattie McKenzie, James Jacobson

Key points

From the paper's conclusion
  • Conclusion The studies verified that test results for the Atellica IM VitDII assay were equivalent between capillary and venous serum, and that vitamin D in capillary samples was stable for 72 hours at 2-8°C and 24 hours at 20-25°C. *In development.

Abstract

Background Capillary blood collection is a convenient, less invasive alternative to venipuncture used in a wide range of clinical and ambulatory settings. Capillary blood produces accurate diagnostic results from small sample volumes and facilitates both point-of-care and core laboratory testing. It is an asset for testing geriatric and pediatric populations, as well as those with difficult venous access. The first study compared test results from adult capillary and venous serum using the Vitamin D Total II (VitDII) assay* on the Atellica IM Analyzer; the second study established capillary sample stability for the assay. Methods Two IRB-approved clinical studies were performed at the Babson Diagnostics site with ambulatory adult subjects enrolled with informed consent. The specimen equivalence study was designed according to CLSI EP35. One hundred eighteen (118) matched set samples were collected: venous serum using Vacutainer SST (comparator) and capillary serum using Microtainer SST. A total of 108 native matched sets were assayed, and 10 contrived matched sets spiked with 25(OH)D3 stock for concentrations across the measuring interval. Of the 118 sets, final analysis was performed on 113 due to loss of replicates including failed venipuncture attempts. Samples were collected on a rolling basis and assayed in duplicate within 4 hours of collection using one reagent lot on one Atellica IM Analyzer. Data were analyzed by Weighted Deming fit regression of the venous comparator (X-axis) versus capillary (Y-axis) samples, utilizing the first replicate from each sample to estimate slope, intercept, and 95% CI for each parameter. Relative repeatability was estimated by pooled SD ratio. For capillary sample stability, four capillary sample sets per subject were collected in Microtainer tubes (one per timepoint) and analyzed using one reagent lot, one analyzer, and five replicates per timepoint. For room temperature (20-25°C) storage, seven sample sets were collected, five remained native, and two were contrived (dose increased with 25(OH)D3), with centrifuged samples stored at ambient temperature between timepoints. For refrigerated (2-8°C) storage, nine samples were collected, seven remained native, with centrifuged samples refrigerated between timepoints. Aggregate sample stability was determined using trend analysis, with allowable drift of =10.0% concentration change, or absolute bias =3.20 ng/mL, whichever was greater. Results Method comparison of capillary versus venous serum (n=113) yielded the following equation: y=1.01x+0.10 (ng/mL) [y=1.01x+0.25 (nmol/L)] at a sample interval of 5.36–117.87 (ng/mL) [13.40–294.68 (nmol/L)], r=0.992, fulfilling requirements. Slope 95% CI was 0.97–1.06 and y-intercept 95% CI was -0.74–0.94 ng/mL [-1.85–2.35 nmol/L]. Full range repeatability SD was 1.818 for capillary serum and 1.681 for venous serum, with ratio 1.082, and ratio 95% CI 0.899–1.302. Vitamin D in capillary serum was stable at 2-8oC for 72 hours and 20-25oC for 24 hours. Conclusion The studies verified that test results for the Atellica IM VitDII assay were equivalent between capillary and venous serum, and that vitamin D in capillary samples was stable for 72 hours at 2-8°C and 24 hours at 20-25°C. *In development. Not available for sale. Future availability cannot be guaranteed.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers