Accuracy and Reliability of Pulse Oxygen Saturation Measured Using Wrist-Worn Oximeters.
Oscar Moreigneaux, Maxime Ceugniez, Francois J Lhuissier, Laura Denis, Julia Bourse, Thomas Gille, Eric Hermand
Universite Sorbonne Paris Nord, Département STAPS, Île-de-France, France, Bobigny. · Université du Littoral Côte d'Opale, ULR 7369 URePSSS, Hauts-de-France, France, Dunkirk.
Key points
From the paper's conclusion- This study aimed to evaluate the accuracy of pulse oxygen saturation measured using wrist-worn devices (Garmin Forerunner 965, Polar Vantage 3 and Checkme O2), from sea level to high altitude.
Abstract
This study aimed to evaluate the accuracy of pulse oxygen saturation measured using wrist-worn devices (Garmin Forerunner 965, Polar Vantage 3 and Checkme O2), from sea level to high altitude. Twenty-six healthy subjects underwent five tests under normoxia and hypoxia conditions (simulated altitudes from 3,000 to 5,500 m), consisting of a 5-minute rest phase, followed by 5 minutes of mild exercise. Bland-Altman parameters, absolute error (around±3% of the criterion), mean absolute percentage error, and the lower bounds of intraclass correlation coefficient and concordance correlation coefficient were calculated. Device-measured pulse oxygen saturation values were different from the criterion values (from p<0.01 to p<0.001) and biases were different from 0 (positive for Garmin and CheckmeO2 and negative for Polar); absolute error rates exceeded 50% for all devices. Intraclass correlation coefficient criteria for Polar and Garmin were poor but good for Checkme O2, whereas concordance correlation coefficient were poor for all devices, overall and at every altitude. All devices failed to provide trustworthy pulse oxygen saturation values at both sea level and altitude, where a pronounced arterial oxygen desaturation could lead to acute mountain sickness in hypoxia-sensitive subjects, and its life-threatening complications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Agreement of consumer wearable pulse oximeters across activity conditions and skin pigmentation.Frontiers in digital health · 2026
- Pulse Oximetry Versus Arterial Co-Oximetry During Exercise: A Systematic Review of Agreement and Sensor Technology Reporting Gaps.Biosensors · 2026
- Performance validation of a new finger-root-adapted ring-shaped oximeter sensor during stepwise steady-state inspired hypoxia.Journal of clinical monitoring and computing · 2026
- The impact of skin tone on performance of pulse oximeters used by NHS England COVID Oximetry @home scheme: measurement and diagnostic accuracy study.BMJ (Clinical research ed.) · 2026
- Accuracy of Optical Heart Rate Measurements for 10 Commercial Wearables in Different Climate Conditions and Activities: Instrument Validation StudyJMIR Formative Research · 2026
- Pulse Oximetry and Arterial Blood Gas Agreement in an Outpatient Black Lung Clinic: Implications for Clinical Assessment and Decision Making.Journal of primary care & community health · 2026