Development of electrochemiluminescence immunochromatographic test strips for dengue virus.

Qinyi Zhang, Yao Qin, Xihan Wang, Liting Xiao, Yanshan Mai, Jie Wu, Hongbin Zhang, Ling Li

School of Public Health, Guangdong Medical University, Zhanjiang, China. · Department of Basic Medical Research, General Hospital of Southern Theater Command, Guangzhou, China.

Key points

From the paper's conclusion
  • Then, cyclic voltammetry scanning was carried out in the test zone at a rate of 0.1 V/s for electrochemiluminescence (ECL) detection.

Abstract

INTRODUCTION: Dengue virus (DENV) infection poses a serious threat to global public health. Rapid and sensitive point-of-care diagnostic methods are urgently needed for early detection and epidemic control. METHODS: The dengue virus monoclonal antibody NS1-McAb1# was labeled as the labeled antibody and sprayed onto the binding pad using ruthenium tripyridine as the signal probe, and the dengue virus monoclonal antibody DN-REAB-G2-006 was fixed on the detection pad as the capture antibody. To mitigate potential interference from other substances in the sample, the sample pad and conjugate pad was pretreated with a solution containing heterophilic blocking reagent, and Elecsys buffer was selected as the optimal co-reactant. After dropping the sample onto the sample pad, chromatography and immunoreaction were performed sequentially. Then, cyclic voltammetry scanning was carried out in the test zone at a rate of 0.1 V/s for electrochemiluminescence (ECL) detection. A novel detection method was established by optimizing various parameters. RESULT: The ruthenium-labeled antibody was prepared by reacting Ru-NHS ester with NS1-McAb1# at a molar ratio of 50:1 for 10 h. Elecsys Buffer was selected as the co-reactant, with a sample-to-buffer volume ratio of 2:1. Whatman AE 100 membrane was used as the detection pad with a length set to 15 mm. The concentration ratio of labeled antibody to capture antibody was 1:2. The chromatographic time after sample loading was 10 min. Under these optimized conditions, the detection method achieved a minimum detection limit of 20.76 pg/mL, exhibited good specificity, with an intra-assay repeatability coefficient of variation of 4%, a storage stability coefficient of variation of 0.83%, and an inter-assay stability coefficient of variation of 0.81%. Clinical sample validation showed a Kappa value of 0.625 compared with nucleic acid testing results. DISCUSSION: The developed ECLIA test strip provides a portable, rapid, and sensitive detection method for DENV, with potential applications in on-site testing across multiple scenarios. Although preliminary clinical evaluation showed acceptable concordance with nucleic acid testing, further large-scale validation using diverse clinical specimens is necessary to fully establish its diagnostic utility in real-world settings.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers