A Paper-Based Electrochemiluminescence Lateral Flow Immunoassay Using Ruthenium Complex Labeling for the Detection of Naja Naja Venom.
Yicai Gan, Shichao Yuan, Wei Chen, Hongbin Zhang, Bo Xu
The First School of Clinical Medicine, Southern Medical University. · Laboratory of Basic Medical Sciences, General Hospital of Southern Theater Command of PLA.
Key points
From the paper's conclusion- These findings support the analytical feasibility of the competitive paper-based ECL platform and provide a basis for further matrix, cross-reactivity, stability, and clinical-sample validation.
Abstract
An indirect competitive paper-based electrochemiluminescence (ECL) lateral flow immunoassay was developed for quantitative detection of Naja naja venom. Equine anti-Naja naja venom IgG served as the primary recognition antibody, ruthenium-complex-labeled rabbit anti-equine IgG as the ECL reporter and immobilized three-finger toxin (3FTX) as the competitive antigen. Ruthenium labeling conditions, the co-reactant system, detection membrane, membrane length, and the labeled-secondary-antibody-to-3FTX ratio were optimized. The selected conditions were a Ru-NHS (ruthenium-complex N-hydroxysuccinimide) ester-to-secondary-antibody molar ratio of 160:1 at 25 °C, Elecsys buffer as the co-reactant, Whatman AE 100 as a 1.5 cm detection membrane, and a 3000:1 secondary-antibody-to-3FTX concentration ratio. The whole-venom calibration series was fitted with a decreasing four-parameter logistic model (R2 = 0.9980). Blank-threshold analysis followed by six-run low-concentration verification supported a practical limit of detection of 32 ng/mL. Quantitative verification identified a limit of quantitation (LOQ) of 300 ng/mL and a validated quantitative range of 300-10,000 ng/mL. The within-batch coefficient of variation (CV) was 3.46%, and the coefficient of variation among three batch means was 1.63%. At 10,000 ng/mL, Naja naja venom produced a lower ECL response than Deinagkistrodon acutus venom, Bungarus multicinctus venom, serum, and phosphate-buffered saline controls. These findings support the analytical feasibility of the competitive paper-based ECL platform and provide a basis for further matrix, cross-reactivity, stability, and clinical-sample validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development of electrochemiluminescence immunochromatographic test strips for dengue virus.Frontiers in bioengineering and biotechnology · 2026
- Development of a lateral flow assay as a point-of-care diagnostic device for the detection of poisoning with cholinesterase-inhibiting compounds.Toxicology letters · 2026
- Nanobody-based lateral flow assays for the detection of α-bungarotoxin in Bungarus snake venomsScientific Reports · 2026
- Venom detectives: a sandwich ELISA platform for precision diagnosis of snakebite envenoming.Frontiers in toxicology · 2026
- Establish a visual loop-mediated isothermal amplification (LAMP) method for rapid detection avian colibacillosis.Frontiers in veterinary science · 2026
- Nanoconfinement catalysis-enhanced electrochemiluminescence aptasensor for the sensitive detection of matrix metalloproteinase-9RSC Advances · 2026