White Blood Cell Counting in Severely Leukopenic Samples on a Sysmex XN Hematology Analyzer: Analytical Performance Evaluation of Whole Blood Mode and Low White Blood Cell Mode.
Jongho Yi, Hanah Kim, Gun-Hyuk Lee, Hyejung Lee, SeungWan Kim, Mina Hur
Department of Laboratory Medicine, Konkuk University School of Medicine, Seoul, Korea. · Department of Laboratory Medicine and Genetics, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea.
Key points
From the paper's conclusion- This is the first study to evaluate the analytical performance of WBC counting in WB and LW modes on XN in samples with WBC <0.50×109/L according to guidelines. The two modes demonstrated reliable analytical performance. As precision decreased at WBC <0.2×109/L and discrepancies between the two modes were rare, careful attention is recommended.
Abstract
BACKGROUND: Sysmex XN-9100 (XN) measures WBC counts in whole blood (WB) mode and automatically switches to low WBC (LW) mode for leukopenic samples (WBC <0.50×109/L). However, the analytical performance of LW mode in such samples has not been evaluated according to standard guidelines. We evaluated the analytical performance of the two modes in samples with WBC <0.50×109/L. METHODS: Precision of WBC quantification in the two modes was evaluated using four samples according to the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) criterion (CV% 5.6). Linearity was evaluated at five concentrations. The two modes were compared in 650 leukopenic samples. Two samples showing discrepancies between the two modes were reviewed. RESULTS: WBC counting precision met the EFLM criterion for both modes, except in a sample with WBC <0.2×109/L. Linearity was acceptable and within the allowable deviation of ±4.8% for both modes. Very high correlation (r=0.99), minimal mean difference (0.00×109/L), and almost perfect agreement (κ=0.90) were noted. Discrepant samples were interpreted as clinically consistent in WB and LW modes in one case each. CONCLUSIONS: This is the first study to evaluate the analytical performance of WBC counting in WB and LW modes on XN in samples with WBC <0.50×109/L according to guidelines. The two modes demonstrated reliable analytical performance. As precision decreased at WBC <0.2×109/L and discrepancies between the two modes were rare, careful attention is recommended.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Assessing intra- and inter-analyser imprecision in automated hematological laboratories.Journal of medical biochemistry · 2026
- Real-World Evaluation of the Sysmex XN-9100 WDF-WPC Workflow: Diagnostic Performance, False-Negative Burden, and Potential Optimization of Smear Review Workload.Diagnostics (Basel, Switzerland) · 2026
- Comparison of the HemoScreen Point-Of-Care Haematology Analyser With Sysmex XN, Beckman Coulter DxH 560 and HORIBA Yumizen H550 in Patients With Abnormal Blood Counts.International journal of laboratory hematology · 2026
- A-226 Benchtop Sysmex XN-450: Evaluation for Use in Primary Healthcare SettingClinical Chemistry · 2026
- Point-of-Care White Blood Cell Differential Testing in Pediatric Inpatients (2 Months-17 Years): How Does HemoScreen Perform?International journal of laboratory hematology · 2026
- Optimization of Leukocytic Flags in the Sysmex-XN Series With the Help of Flow Cytometric Differential (FlowDiff) Can Significantly Optimize Laboratory Efficiency.International journal of laboratory hematology · 2026