Optimization of Leukocytic Flags in the Sysmex-XN Series With the Help of Flow Cytometric Differential (FlowDiff) Can Significantly Optimize Laboratory Efficiency.
Ninna Hahn Tougaard, Snorre Kristian Vesten Pedersen, Anders Abildgaard, Anne Winther-Larsen, Konstantinos Dimopoulos
Department of Clinical Biochemistry, Bispebjerg and Frederiksberg Hospital, Copenhagen, Denmark. · Department of Clinical Biochemistry, Aarhus University Hospital, Aarhus, Denmark.
Key points
From the paper's conclusion- The count flags ‘eosinophilia’, ‘basophilia’ and ‘monocytopenia’ were deactivated, as the results from Sysmex XN were comparable to the flow cytometric reference method.
Abstract
OBJECTIVES: The aim of this study was to adjust the default white blood cell (WBC) differential count flag settings on the Sysmex XN-9100 haematology analyser to reduce unnecessary additional leukocyte differential counts without compromising diagnostic sensitivity. METHODS: Data from 4554 routine blood samples flagged on Sysmex XN-9100 and corresponding additional leukocyte differential counts performed on the Sysmex XF-1600 flow cytometer were used to optimise the WBC flagging rules. RESULTS: A total of 1148 samples had scattergram-related flags, which are non-adjustable. Optimal cut points for Q-flags were established based on Youden's Index for the remaining 3406 samples, while the diagnostic sensitivity of the new flagging rules for acute leukaemia was assessed in a validation cohort. The optimal cut points for Q-flags were 120 for 'Atypical Lymphocytes', 170 for 'Abnormal Lymphocytes' and 160 for 'Blasts', while the Q-flag 'Left shift' offered no diagnostic value in our cohort. The count flags 'eosinophilia', 'basophilia' and 'monocytopenia' were deactivated, as the results from Sysmex XN were comparable to the flow cytometric reference method. By implementing the new flagging rules, 1709 (37.5%) of the samples would have been processed without need for additional leukocyte differential count. When evaluated in a separate validation cohort of 90 samples from patients with newly diagnosed acute leukaemia, the new rules showed similar diagnostic sensitivity to the default flagging system. CONCLUSION: Optimisation of Sysmex flagging rules for leukocyte differential counts was feasible and significantly reduced unnecessary reflex testing, while maintaining analytical accuracy and diagnostic safety.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Real-World Evaluation of the Sysmex XN-9100 WDF-WPC Workflow: Diagnostic Performance, False-Negative Burden, and Potential Optimization of Smear Review Workload.Diagnostics (Basel, Switzerland) · 2026
- Point-of-Care White Blood Cell Differential Testing in Pediatric Inpatients (2 Months-17 Years): How Does HemoScreen Perform?International journal of laboratory hematology · 2026
- Utility of WBC Scatterplots and Suspect Flags Generated by Beckman Coulter LH-750 Hematology Analyzer in the Characterization of Leukemias and Related Hematological Malignancies.Indian journal of hematology & blood transfusion : an official journal of Indian Society of Hematology and Blood Transfusion · 2026
- Beyond Analytical Agreement: Integrating Analytical Performance and Algorithm-Assisted Morphological Interpretation in a Comparative Evaluation of the Mindray BC-7800/BC-7900 and Sysmex XN-9100 Hematology Platforms.Diagnostics (Basel, Switzerland) · 2026
- B-075 Can DIFF Scattergrams Predict Acute Leukemia Phenotype? A Standardized Interpretation Approach Using Mindray BC-6800PlusClinical Chemistry · 2026
- Analytical Performance of the Sysmex XF ‐1600 Flow Cytometer in Routine Clinical PracticeInternational Journal of Laboratory Hematology · 2026