Comparison of the HemoScreen Point-Of-Care Haematology Analyser With Sysmex XN, Beckman Coulter DxH 560 and HORIBA Yumizen H550 in Patients With Abnormal Blood Counts.

Katarína Hradská, Martin Pulcer, Michaela Skořupová, Tereza Popková, Michal Kaščák, Jana Mihályová, Ivo Demel, Barbora Dluhošová, Kateřina Benková, Juraj Ďuraš, Soňa Kolářová, Veronika Hanzelková, Simona Pulcerová, Roman Hájek

一句话结论 · In one sentence

HemoScreen demonstrated good analytical agreement with all evaluated haematology analysers and may represent a reliable complementary POCT tool for routine haematological monitoring, including patients with abnormal blood counts.

原始摘要(英文原文)· Original abstract
INTRODUCTION: HemoScreen is a portable point-of-care testing (POCT) haematology analyser capable of measuring 20 complete blood count (CBC) parameters, including a 5-part leukocyte differential, from venous and capillary blood. This study compared selected CBC and white blood cell (WBC) differential parameters obtained using HemoScreen with those from three conventional haematology analysers: Sysmex XN, Beckman Coulter DxH 560, and HORIBA Yumizen H550 in abnormal samples. METHODS: Venous and capillary blood was collected from 137 patients with haematological disorders and healthy donors. We evaluated WBC, absolute neutrophil count (ANC), haemoglobin (HGB), and platelets (PLT). Statistical analysis included Passing-Bablok regression analysis, Cusum test, Wilcoxon signed-rank test, and Hodges-Lehmann median difference analysis. RESULTS: Comparison of venous and capillary blood measurements using HemoScreen demonstrated no significant deviation from linearity for any evaluated parameter according to the Cusum test (p = 0.14-0.98). Although statistically significant differences were observed for ANC (p = 0.0201) and PLT (p = 0.0026), the median differences remained small and clinically irrelevant. Comparison of venous blood measurements between HemoScreen and conventional analysers demonstrated strong analytical agreement without significant deviation from linearity. Statistically significant differences were observed in several comparisons, particularly for PLT and HGB measurements; however, the approximate relative bias remained within external quality assessment limits in most comparisons, with greater variability for PLT in the ≤ 50 × 109/L subgroup. CONCLUSIONS: HemoScreen demonstrated good analytical agreement with all evaluated haematology analysers and may represent a reliable complementary POCT tool for routine haematological monitoring, including patients with abnormal blood counts.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Comparison of the HemoScreen Point-Of-Care Haematology Analyser With Sysmex XN, Beckman Coulter DxH 560 and HORIBA Yumizen H550 in Patients With Abnormal Blood Counts. — 科研速览 Science Skim