Laboratory evaluation of the MedMira Multiplo Complete Syphilis (TP/nTP) Antibody Test using paired serum and simulated whole-blood specimens.
Kathleen Whyte, Meika Ei Richmond, Michelle Shuel, William Hoang, Raymond Sw Tsang, Vanessa Tran
Laboratory for Syphilis Diagnostics, National Microbiology Laboratory Branch, Public Health Agency of Canada, Winnipeg, Manitoba, Canada. · Public Health Ontario, Toronto, Ontario, Canada.
Key points
From the paper's conclusion- Further evaluation in the field is necessary to determine the performance and acceptability of this device in clinical settings.
Abstract
BACKGROUND: Syphilis cases have been on the rise in Canada, and rapid point-of-care test (POCT) options may aid in timely diagnosis and treatment, thereby preventing further spread of infection. This study provides a laboratory evaluation of the performance of the MedMira Multiplo Complete Syphilis (TP/nTP) Antibody Test in comparison with the current conventional testing methods. METHODS: Two operators independently performed testing in a blinded manner using 130 paired serum and simulated blood specimens, with the results compared with the corresponding known conventional syphilis serology test results. RESULTS: Overall, among specimens known to be CMIA-positive and RPR-reactive, with a titre ≥1:8, the sensitivity and specificity of the POCT when tested with sera were 97.5% and 70%, respectively, for operator 1, and 100% and 61.1%, respectively, for operator 2. With simulated blood specimens and the same criteria, the sensitivity and specificity of the POCT were 95.0% and 82.2%, respectively, for operator 1, and 100.0% and 74.4%, respectively, for operator 2. CONCLUSION: Further evaluation in the field is necessary to determine the performance and acceptability of this device in clinical settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Association of treponemal chemiluminescent immunoassay signal intensity with TPPA reactivity in reverse sequence syphilis screening.Microbiology spectrum · 2026
- Development and evaluation of a rapid duplex qPCR assay for Treponema pallidum using the GeneSoC microfluidic platform.Diagnostic microbiology and infectious disease · 2026
- B-160 Streamlining Syphilis Serology Through an Automated Multiplex Syphilis Total and RPR PanelClinical Chemistry · 2026
- B-190 Increasing Syphilis Positivity and the Importance of Validated Rapid Testing for Early Diagnosis and Disease Control StrategiesClinical Chemistry · 2026
- Interlaboratory comparison of a multiplex bead assay against the Chlamydia trachomatis antigen Pgp3 using two monoclonal antibody clones.PLOS global public health · 2026
- Field Performance and usability of a novel molecular point-of-care assay for Chlamydia trachomatis and Neisseria gonorrhoeae detection.International journal of STD & AIDS · 2026