Interlaboratory comparison of a multiplex bead assay against the Chlamydia trachomatis antigen Pgp3 using two monoclonal antibody clones.
Sarah Gwyn, Karana Wickens, Helena Brazal Monzo, Chris Drakeley, Stephane Pelleau, Michael White, Zachary Reynolds, Henry Kanyi,
Division of Parasitic Diseases and Malaria, National Center for Zoonotic, Waterborne, and Environmental Diseases, Centers for Disease Control and Prevention, Atlanta, Georgia, United States of America. · Department of Immunology, Faculty of Infectious and Tropical Diseases, London School of Hygiene and Tropical Medicine, London, United Kingdom.
Key points
From the paper's conclusion- These data suggest that 1C3 may be the more reliable standard reagent and that the Pgp3 multiplex assays are highly robust when comparing different instruments and assay protocols.
Abstract
The bacterium Chlamydia trachomatis causes three distinct conditions - urogenital chlamydia, the eye disease trachoma, and lymphogranuloma venereum. Use of serological testing for exposure to C. trachomatis for evaluating ocular disease risk factors and population-based surveillance has expanded over the past decade. We developed chimeric monoclonal antibodies to the C. trachomatis antigen Pgp3 as standardized reagents for Pgp3 serological assays. Here we use two anti-Pgp3 monoclonals (1G1 and 1C3) to conduct an interlaboratory comparison of Pgp3 Luminex-based multiplex bead serological assays in 4 different laboratories (each with their own protocol) with 5 operators using 3 different instrument types. Antibodies spotted onto filter paper were stable for two weeks at room temperature. Accuracy and precision were evaluated at 3 concentrations for each monoclonal (high, medium, and low). Accuracy within 20% was achieved for 11/21 conditions for 1G1 and 18/21 conditions for 1C3; the high concentration resulted in the poorest accuracy. Precision was within 10% for all low and medium concentrations and was above 10% for several conditions at the high concentration. The fluorescence values differed between instruments but showed no difference when extrapolated to concentrations using a standard curve. These data suggest that 1C3 may be the more reliable standard reagent and that the Pgp3 multiplex assays are highly robust when comparing different instruments and assay protocols.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A multiplex real-time PCR subtyping assay combining chromosomal (pmpH) and plasmid (pgp3) markers for differentiation of LGV and non-LGV Chlamydia trachomatis.Microbiology spectrum · 2026
- Paired molecular characterization of repeat Chlamydia trachomatis infection among pregnant women in East London, South AfricamedRxiv · 2026
- Performance evaluation and application of a multiplex PCR capillary electrophoresis method for detecting nucleic acids of seven sexually transmitted pathogensFrontiers in Cellular and Infection Microbiology · 2026
- Rapid, low-cost colorimetric capillary-driven immunoassay for screening of bacterial STIs: Chlamydia, syphilis, and gonorrhea.Biosensors & bioelectronics · 2026
- Development of a recombinant MOMP-based indirect ELISA for the serodiagnosis of caprine chlamydiasis.Tropical animal health and production · 2026
- Characterisation and manufacture of a Neisseria gonorrhoeae challenge agent for use in an oropharyngeal controlled human infection modelbioRxiv · 2026