Long-Term in vivo Positional Stability of a Suprachoroidal Telemetric Intraocular Pressure Sensor Implant
Colya N. Englisch, Philip Wakili, André M. Trouvain, Wouter J. van Drunen, H. Burkhard Dick, Kaweh Mansouri, Esther M. Hoffmann, Marc J. Mackert,
Knappschaftsklinikum Saar · Implandata Ophthalmic Products (Germany)
Key points
From the paper's conclusion- Background: The aim of this study was to evaluate the long-term in vivo positional stability of a novel suprachoroidal telemetric intraocular pressure (IOP) sensor implant.
Abstract
Background: The aim of this study was to evaluate the long-term in vivo positional stability of a novel suprachoroidal telemetric intraocular pressure (IOP) sensor implant. Methods: In this prospective, open-label, multicenter interventional study, 24 patients undergoing nonpenetrating glaucoma surgery with simultaneous implantation of the suprachoroidal EyeMate-SC IOP sensor were considered (Implandata Ophthalmic Products GmbH, Hannover, Germany). Overall, 11 patients were excluded owing to incomplete data. Of the remaining patients, 11 were diagnosed with primary open-angle glaucoma, 1 with pseudoexfoliative glaucoma, and 1 with uveitic glaucoma. The sensor position was measured intraoperatively via surgical video analysis and postoperatively via swept-source anterior segment optical coherence tomography (Anterion, Heidelberg Engineering GmbH, Heidelberg, Germany) with reference to the anatomical limbus. Postoperative assessments were conducted at 10, 30, 90, 180, 270, 360, 540, 720, 900, and 1080 days. Results: Unilateral implantation was successful in all 24 cases. No dislocations occurred. In one patient, partial suprachoroidal positioning was observed intraoperatively, with subsequent autonomous repositioning into full suprachoroidal placement within weeks. Excluding this case, the absolute implant position was 1.6 ± 0.8 mm at day 0, 1.7 ± 1.0 mm at day 360, 1.9 ± 1.8 mm at day 720, and 2.1 ± 0.9 mm at day 1080. Notably, the relationship between ΔIOP (Goldmann applanation tonometry - EyeMate-SC reading) and absolute implant position remained nonsignificant, regardless of whether the patient with the initially incomplete implantation was included (r = -0.08, 95% CI: -0.32 to 0.38) or excluded (r = -0.11, 95% CI: -0.37 to 0.36). Conclusion: This study indicates high long-term in vivo positional stability of the EyeMate-SC IOP sensor. The observed variations were within measurement uncertainty. This supports the sensor's potential for safe and continuous IOP monitoring in glaucoma care.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Intraoperative OCT observation of incision-related descemet membrane detachment during implantable collamer lens implantation.Frontiers in medicine · 2026
- Review on Intraocular Pressure-Sensing Systems from Fabrication to Applications.ACS sensors · 2026
- Outcomes of Carlevale-type scleral-fixated intraocular lenses as a stable platform in silicone oil-filled eyes: a multicenter studyScientific Reports · 2026
- Selective Laser Trabeculoplasty in Open-Angle Glaucoma and Ocular Hypertension: A Narrative Review of Current Evidence.Journal of clinical medicine · 2026
- First-in-Human Implantation of an Add-On Dual Intraocular System Combining a Standard IOL and the SING-IMT™ Telescope: Optical Simulation, Preclinical Validation, and Clinical Feasibility.Retina (Philadelphia, Pa.) · 2026
- Influencing Factors of Positional Stability of Implantable Collamer Lenses in Terms of Rotation, Decentration, and Tilt in Patients With Myopia.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026