First-in-Human Implantation of an Add-On Dual Intraocular System Combining a Standard IOL and the SING-IMT™ Telescope: Optical Simulation, Preclinical Validation, and Clinical Feasibility.
Stanislao Rizzo, Paola Sasso, Irene Nepita, Giuseppe Francione, Carmela G Caputo, Mariantonia Ferrara, Alessandro Feo, Alfonso Savastano,
Ophthalmology Department, "Fondazione Policlinico Universitario A. Gemelli, IRCCS", Rome, Italy. · CNR-INO, Istituto Nazionale di Ottica, BEAM Lab, Florence, Italy.
Key points
From the paper's conclusion- This integrated evaluation supports the feasibility and short-term safety of sulcus-based SING-IMT implantation as a minimally invasive intraocular magnification strategy for selected pseudophakic patients with advanced AMD.
Abstract
PURPOSE: To evaluate a dual-implant system combining a capsular-bag monofocal intraocular lens with a sulcus-implanted small-incision new-generation implantable miniature telescope (SING-IMT). METHODS: Optical simulations quantified retinal spot size, modulation transfer function (MTF), and tolerance to misalignment. Preclinical porcine experiments evaluated surgical feasibility, anatomical stability, and early safety. First-in-human (FiH) implantation assessed surgical outcomes, anterior segment behaviour using ultrasound biomicroscopy and optical coherence tomography, functional vision, and safety over 6 months. RESULTS: Optical modelling demonstrated that the dual-implant configuration maintained retinal image quality within the rehabilitative regime for advanced AMD, with root-mean-square retinal spot radii of 18.7 µm (0°) and 20.4 µm (2.5°), below the estimated retinal sampling limit (∼49 µm), consistent with an expected visual acuity improvement of approximately 2 Snellen lines. Decentration up to 0.5 mm was optically well tolerated, whereas tilt produced rapid mid-frequency contrast loss. In vivo porcine studies confirmed stable sulcus seating without disruption of the capsular-bag IOL or anterior segment structures. FiH implantation was completed without intraoperative complications. At 6 months, distance visual acuity improved from 20/200 to 20/125, near acuity from 56 to 15 Colenbrander units, and reading speed from non-assessable preoperatively to 20 words/min, with no device-related adverse events. CONCLUSION: This integrated evaluation supports the feasibility and short-term safety of sulcus-based SING-IMT implantation as a minimally invasive intraocular magnification strategy for selected pseudophakic patients with advanced AMD.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Clinical and simulated impact of intraocular lens tilt and decentration based on real-world data and optical simulationScientific Reports · 2025
- Extended Depth-of-Focus Intraocular Lens Based on Aberration-Neutral Technique.Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists) · 2026
- Binocular intermediate vision enhancement without compromising optical quality using a SwissPhob enhanced monofocal intraocular lens.Indian journal of ophthalmology · 2026
- Clinical Outcomes Following Bilateral Implantation of a Next-Generation Intraocular Lens in Patients with Primary Open Angle Glaucoma.Clinical ophthalmology (Auckland, N.Z.) · 2026
- Biconcave freeform implantable contact lenses: wide-field design and peripheral aberration optimization.Optics express · 2026
- Clinical outcomes of a new presbyopia-correcting intraocular lens with spiral optic.Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie · 2026