The second assessment of the commutability of candidate reference materials for the harmonization of blood neurofilament light chain measurements.
Ulf Andreasson, Vincent Delatour, Aleksandra Maceski, Johan Gobom, Burak Arslan, Elena Camporesi, Chiara Giangrande, Guy Auclair,
Department of Psychiatry and Neurochemistry, The Sahlgrenska Academy at the University of Gothenburg, Mölndal, Sweden. · Laboratoire National de Métrologie et d'Essais, Paris, France.
Key points
From the paper's conclusion- Spiking serum pools with CSF appears to be an effective strategy for consistently producing commutable RMs for NfL measurement. Although the relatively high proportion of indeterminate results indicates a need for improved MP precision or additional replicate measurements, the study supports continued development of SI-traceable reference materials to facilitate standardization of NfL measurements in clinical practice.
Abstract
OBJECTIVES: Blood neurofilament light (NfL) is a sensitive biomarker of neuroaxonal injury, but substantial differences in absolute concentrations across analytical platforms underscore the need for assay standardization. This study aimed to evaluate the commutability of candidate reference materials (RMs) generated by spiking serum pools with cerebrospinal fluid (CSF) to support global assay harmonization. METHODS: The study included 45 individual clinical serum samples and 30 candidate RMs. The RMs comprised three serum pools, each combined with each of three CSF pools at three concentration levels, in addition to the three unspiked serum pools. Commercial NfL measurement procedures (MPs) from six manufacturers (Roche, Fujirebio, Meso Scale Discovery, Bio-Techne, UmanDiagnostics, and Quanterix) were evaluated. Commutability was assessed using a difference in bias approach. RESULTS: Spearman correlations between MPs were excellent (ρ=0.96-0.99), although regression slopes differed substantially, with differences up to a 46-fold. Across the 450 RM assessments, 62 % were classified as commutable, 37 % as indeterminate, and 0.66 % as non-commutable. Approximately two-thirds of the uncertainty in the difference in bias was attributable to MP imprecision rather than sample specific effects, suggesting that standardization of NfL measurements is feasible. CONCLUSIONS: Spiking serum pools with CSF appears to be an effective strategy for consistently producing commutable RMs for NfL measurement. Although the relatively high proportion of indeterminate results indicates a need for improved MP precision or additional replicate measurements, the study supports continued development of SI-traceable reference materials to facilitate standardization of NfL measurements in clinical practice.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Highly Reproducible Quantification of Serum Neurofilament Light Chain in a Multi-Center Study.European journal of neurology · 2026
- Real-world Roche Elecsys® and Fujirebio Lumipulse® inter-platform comparability and clinical validation of plasma neurofilament light chain in multiple sclerosis.Journal of neurology · 2026
- Plasma and CSF neurofilament light chain measured by Simoa and Lumipulse: an inter-platform comparison across neurological disorders.Clinical chemistry and laboratory medicine · 2026
- Evaluating feasibility of ferritin assay harmonisation with commutable proof-of-concept reference materials and EQA-based monitoring.Clinical chemistry and laboratory medicine · 2026
- Comparative analysis of three platforms for serum NfL quantification in healthy controls and MS patientsClinical Chemistry and Laboratory Medicine (CCLM) · 2025
- B-035 Validation of Neurofilament Light Chain as laboratory developed test using electrochemiluminescence assay in serum and plasmaClinical Chemistry · 2026