Real-world Roche Elecsys® and Fujirebio Lumipulse® inter-platform comparability and clinical validation of plasma neurofilament light chain in multiple sclerosis.
Vincenzo Criscuolo, Valerio Nicolella, Carmela Polito, Rosa Sirica, Mariano Fiorenza, Domenica Sorvillo, Ilaria Delle Cave, Antonio Carotenuto,
Department of Neuroscience, Reproductive Science and Odontostomatology, Federico II University of Naples, Naples, Italy. · Department of Translational Medical Sciences, Federico II University of Naples, Naples, Italy.
Key points
From the paper's conclusion- Elecsys® and Lumipulse® are not interchangeable on the raw pg/mL scale though conversion-based approaches can improve cross-platform alignment within the observed analytical range. Elecsys® pNfL retained expected clinical associations, supporting routine MS use. Findings support cautious conversion during platform transitions in clinics.
Abstract
BACKGROUND: Neurofilament light chain (NfL) reflects neuro-axonal injury in multiple sclerosis (MS), but method-dependent bias limits interchangeability across automated platforms. We assessed agreement between Roche Elecsys® and Fujirebio Lumipulse®, evaluated Simoa-referenced harmonization, validated Elecsys® clinical associations, and derived directional IVD-to-IVD conversion models between Elecsys® and Lumipulse®. METHODS: In this retrospective real-world cohort (N=253), fasting EDTA plasma was analyzed in singlicate on Elecsys® and Lumipulse®. Clinical associations of Elecsys® pNfL were tested using multivariable linear models, including age, sex, disease duration, EDSS, cardiovascular comorbidities, smoking, and disease course, with sensitivity analyses including BMI. RESULTS: On the raw scale, platforms were strongly correlated but showed poor agreement (Coeff 6.91, 95% CI 6.44-7.38; r=0.879; CCC=0.107, 95% CI 0.090-0.123). After conversion of both assay outputs to published Simoa-equivalent values, alignment improved (Coeff 1.03, 95% CI 0.96-1.10). We derived a directional Elecsys®-to-Lumipulse® Passing-Bablok conversion (Lumipulse = 1.97 + 11.171 × Elecsys). Higher Elecsys® pNfL was associated with disability (EDSS: Coeff 0.302, 95% CI 0.122-0.481; with BMI: Coeff 0.335, 95% CI 0.119-0.551) and was lower in treated versus untreated patients. CONCLUSIONS: Elecsys® and Lumipulse® are not interchangeable on the raw pg/mL scale though conversion-based approaches can improve cross-platform alignment within the observed analytical range. Elecsys® pNfL retained expected clinical associations, supporting routine MS use. Findings support cautious conversion during platform transitions in clinics.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative analysis of three platforms for serum NfL quantification in healthy controls and MS patientsClinical Chemistry and Laboratory Medicine (CCLM) · 2025
- Plasma and CSF neurofilament light chain measured by Simoa and Lumipulse: an inter-platform comparison across neurological disorders.Clinical chemistry and laboratory medicine · 2026
- Fresh-to-frozen comparability of serum neurofilament light chain across four immunoassaysScientific Reports · 2026
- Highly Reproducible Quantification of Serum Neurofilament Light Chain in a Multi-Center Study.European journal of neurology · 2026
- Serum Neurofilament Light Chain and GFAP in Amyotrophic Lateral Sclerosis on a Commercial ECLIA Platform.Muscle & nerve · 2026
- Same biomarker, different number: why multiple sclerosis needs assay-aware NfL interpretation.Multiple sclerosis and related disorders · 2026