Plasma and CSF neurofilament light chain measured by Simoa and Lumipulse: an inter-platform comparison across neurological disorders.
Andrea Toja, Virginia Quaresima, Chiara Tolassi, Tommaso Merati, Chiara Trasciatti, Simona G Signorini, Andrea Morotti, Francesco Berinato,
Department of Continuity of Care and Frailty, Neurology Unit, ASST Spedali Civili Hospital, Brescia, Italy. · Clinical Chemistry Laboratory, Diagnostic Department, ASST Spedali Civili di Brescia, Brescia, Italy.
Key points
From the paper's conclusion- Simoa and Lumipulse capture a consistent plasma NfL signal across neurological disorders, but absolute concentrations differ, supporting platform-specific, age-aware interpretation in clinical practice.
Abstract
OBJECTIVES: Blood neurofilament light chain (NfL) is an accessible biomarker of neuroaxonal injury, but clinical implementation requires robust analytical comparability across platforms. We evaluated plasma NfL measured by Simoa and Lumipulse across neurological disorders, including paired CSF analyses to anchor plasma findings to the central compartment. METHODS: We included 383 individuals, comprising healthy controls and patients with neurodegenerative diseases, multiple sclerosis, and ischemic stroke. Plasma NfL was measured in all participants using Simoa and Lumipulse; paired CSF measurements were available in 92 individuals. Lumipulse intermediate precision and between-day repeatability were assessed according to CLSI EP15. Cross-platform agreement was evaluated using Spearman correlation, Passing-Bablok regression, and Bland-Altman analysis. Plasma-CSF associations and age-adjusted diagnostic differences were assessed separately for each platform. RESULTS: Plasma NfL measured by Simoa and Lumipulse showed strong concordance in the whole cohort (ρ=0.90), with similarly strong CSF concordance in paired samples (ρ=0.90). Method-comparison analyses demonstrated constant and proportional bias, with systematically higher absolute plasma NfL values measured by Lumipulse. Plasma and CSF NfL were significantly correlated within each platform (ρ=0.72 for Simoa; ρ=0.77 for Lumipulse). Lumipulse blood and CSF assays showed high precision. Across both platforms, plasma NfL increased with age and was higher in neurological disorders than in healthy controls. CONCLUSIONS: Simoa and Lumipulse capture a consistent plasma NfL signal across neurological disorders, but absolute concentrations differ, supporting platform-specific, age-aware interpretation in clinical practice.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Real-world Roche Elecsys® and Fujirebio Lumipulse® inter-platform comparability and clinical validation of plasma neurofilament light chain in multiple sclerosis.Journal of neurology · 2026
- Comparative analysis of Ella™ and Simoa™ immunoassays for neurofilament light chain quantification in bovine serum and CSFScientific Reports · 2026
- B-152 Evaluation of the Analytical Performance of Neurofilament Light Chain immunoassaysClinical Chemistry · 2026
- Comparative analysis of three platforms for serum NfL quantification in healthy controls and MS patientsClinical Chemistry and Laboratory Medicine (CCLM) · 2025
- Translating neurofilament light chain testing into clinical practice: a multidisciplinary implementation roadmapClinical Chemistry and Laboratory Medicine (CCLM) · 2026
- The second assessment of the commutability of candidate reference materials for the harmonization of blood neurofilament light chain measurements.Clinical chemistry and laboratory medicine · 2026