Highly Reproducible Quantification of Serum Neurofilament Light Chain in a Multi-Center Study.
Niels Kruse, Jonas Franz, Birgit Otte, Leila Husseini, Stefan Nessler, Katja Akgün, Nicole Heubner, Falk Steffen,
Department of Neuropathology, University Medical Center Göttingen, Göttingen, Germany. · Department of Neurology, Multiple Sclerosis Center, Center of Clinical Neuroscience, University Hospital Carl Gustav Carus, University of Technology Dresden, Dresden, Germany.
Key points
From the paper's conclusion- Our findings support the analytical robustness as well as inter-laboratory comparability of sNfL measurements and its use as a serum biomarker in multi-center studies of neurodegeneration.
Abstract
BACKGROUND: Neurofilament light chain measured in serum (sNfL) is an emerging biomarker for neuro-axonal injury, but its clinical use requires robust reference values and demonstration of inter-laboratory comparability. METHODS: This round-robin study aimed to assess age-dependent sNfL using the Quanterix HD-X Analyzer under laboratory conditions that were strictly standardized across all participating sites to exclude technical outliers while providing reference values on an absolute scale. Besides a group of individuals with relapsing-remitting multiple sclerosis (RRMS, n = 24) we measured a cohort of voluntary blood donors (n = 128) to revisit age-related cut-off values under strict laboratory conditions. Measurements were performed in five German laboratories following a standardized protocol, and all reagents and consumables derived from identical kit lots. RESULTS: In line with previous work, we found that sNfL increased with age by about ~0.15 pg/mL per year within the sampled donor age range of 21-60 years. Also, our results demonstrate that measurements obtained across different laboratories exhibited a high degree of comparability with an average coefficient of variation of 11%. Using quantile regression with identical age-dependent thresholds across sites, we observed consistent, exploratory separation of the small MS cohort from healthy donors (mean AUC 0.79, 95% CI 0.68-0.88). CONCLUSION: Our findings support the analytical robustness as well as inter-laboratory comparability of sNfL measurements and its use as a serum biomarker in multi-center studies of neurodegeneration.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Real-world Roche Elecsys® and Fujirebio Lumipulse® inter-platform comparability and clinical validation of plasma neurofilament light chain in multiple sclerosis.Journal of neurology · 2026
- Comparative analysis of three platforms for serum NfL quantification in healthy controls and MS patientsClinical Chemistry and Laboratory Medicine (CCLM) · 2025
- Age-Dependent z Scores and eGFR-Adjusted Reference Ranges for Neurofilament Light: A Practical Approach for Clinical LaboratoriesClinical Chemistry · 2026
- The second assessment of the commutability of candidate reference materials for the harmonization of blood neurofilament light chain measurements.Clinical chemistry and laboratory medicine · 2026
- Fresh-to-frozen comparability of serum neurofilament light chain across four immunoassaysScientific Reports · 2026
- B-035 Validation of Neurofilament Light Chain as laboratory developed test using electrochemiluminescence assay in serum and plasmaClinical Chemistry · 2026