B-152 Evaluation of the Analytical Performance of Neurofilament Light Chain immunoassays

E Jianfei, Kang Wang, Guoping Wu, Yu Xu, Liangliang Zhang, Renlin Li, Junxi Tian

Key points

From the paper's conclusion
  • The results indicated that reference values were age-dependent.

Abstract

Background Accurate identification of axonal injury is critical for the management of neurodegenerative diseases. Neurofilament light chain (NfL) is a neuron-specific cytoskeletal protein released into cerebrospinal fluid (CSF) and blood following axonal injury. Due to its ability to quantify neuronal damage in blood, NfL is currently applied in the monitoring of various neurological diseases and the assessment of therapeutic efficacy. Detection of NfL provides a novel blood-based biomarker for neurological diseases and overcomes the limitations of invasive techniques, such as CSF testing. This study presents the analytical evaluation of a quantitative NfL immunoassay manufactured by Maccura Biotech Co., Ltd. (Chengdu, China) for use with human serum or plasma. Methods The NfL immunoassay processed on the Maccura Automatic Chemiluminescence Analyzer was evaluated according to CLSI guidelines. Analytical sensitivity, including the limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ), was evaluated in accordance with CLSI EP17-A2 using blank and low-concentration samples. LoQ was defined as the lowest concentration meeting predefined imprecision criteria (CV less than 20%). Precision studies were performed per CLSI EP05-A3 using quality control samples and patient samples with different concentrations, measured in duplicate, two runs per day over 20 days. A reference interval study was performed based on guidance from CLSI C28-A3c28 with a general population from healthy individuals (over 18 years old). Method comparison studies evaluated the Simoa assay and Maccura assay in accordance with CLSI EP09-A3. Passing-Bablok regression analysis was applied to clinical samples spanning a wide concentration range. Clinical performance was evaluated in 75 participants, including 47 in the healthy control (HC) group and 28 in the Alzheimer’s disease (AD) group. Results The limits of blank, detection and quantitation were less than 5.0 pg/mL, 10.0 pg/mL and 15.0 pg/mL, respectively. Repeatability and within-laboratory coefficients of variation were 3.05% and 4.26% for QC 1 (93.9 pg/mL); 2.66% and 4.09% for QC 2 (2296.1 pg/mL); 2.38% and 3.36% for Human Sample 1 (75.4 pg/mL); and 2.69% and 3.79% for Human Sample 2 (1052.3 pg/mL). The reference interval study included a total of 300 individuals. The results indicated that reference values were age-dependent. For individuals under 60 years of age, the reference value (95th percentile) was 40.5 pg/mL (95% CI: 39.3 - 46.7 pg/mL, n=150), for those aged 60 years or older, it was 70.1 pg/mL (95% CI: 67.8 - 72.0 pg/mL, n=150). A total of 40 samples were analyzed using both the Maccura assay and the Simoa NfL assay. The Passing-Bablok regression equation was y = 2.882 + 1.566x, and the Pearson correlation coefficient was 0.992. The NfL concentrations in the HC and AD groups were 15.70 ± 4.309 pg/mL and 48.27 ± 27.43 pg/mL, p<0.0001. Conclusion In summary, the Maccura NfL immunoassay demonstrates high analytical sensitivity and robust precision across the clinically relevant measuring range, and shows strong correlation with Simoa. Furthermore, it enables clinically meaningful discrimination between Alzheimer’s disease patients form the healthy controls.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers