Validation of synthetic M-protein materials for external quality assessment: homogeneity, commutability, and stability.
Moniek M Bioch, Theo Rispens, Corrie M de Kat Angelino, Anke J Ten Haaken-Meijer, Miranda van Berkel, Marc H M Thelen, Joannes F M Jacobs
SKML, Foundation for Quality Assessment in Medical Laboratory Diagnostics, Nijmegen, The Netherlands. · Sanquin Research, Amsterdam, The Netherlands.
Key points
From the paper's conclusion- Due to their proven homogeneity, commutability, and stability, targeted synthetic M-protein EQA materials are suitable for use in M-protein EQA schemes. This allows for the assessment of laboratory performances based on target values, and opens the way for harmonisation and exchangeability of M-protein diagnostic methods. Additionally, it provides a new source for the production of M-protein EQA materials, alleviating the demand for large volumes of patient material needed to organise M-protein EQA schemes.
Abstract
OBJECTIVES: Accurate quantification of M-proteins is important for clinical management of monoclonal gammopathies. Reliable assessment of trueness on M-protein quantification in External Quality Assessment (EQA) schemes is difficult due to the lack of reference methods and materials for target value assignment. We propose a strategy to produce targeted synthetic M-protein EQA materials by spiking known amounts of human(ised) therapeutic monoclonal antibodies (t-mAb) into null serum. This paper validates homogeneity, commutability, and stability of these materials. METHODS: Homogeneity was assessed on two aliquoted synthetic M-protein EQA materials comparing within-vial variation with between-vial variation for sodium, total protein, IgG and M-protein quantification. Commutability was assessed by comparing between-method differences in 51 patient-derived and nine synthetic M-protein EQA materials among various M-protein quantification methods. Stability was checked by reanalysis of two previously distributed synthetic M-protein EQA materials. RESULTS: Within- and between-vial variation were similar in both samples, and therefore homogeneity criteria were met. Commutability was observed for eight out of nine samples, whereas one hypergammaglobulinemic sample was non-commutable. Long-term storage stability at -80 °C was found acceptable in both samples. CONCLUSIONS: Due to their proven homogeneity, commutability, and stability, targeted synthetic M-protein EQA materials are suitable for use in M-protein EQA schemes. This allows for the assessment of laboratory performances based on target values, and opens the way for harmonisation and exchangeability of M-protein diagnostic methods. Additionally, it provides a new source for the production of M-protein EQA materials, alleviating the demand for large volumes of patient material needed to organise M-protein EQA schemes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Strengthening the interpretation of longitudinal EQA data through the sporadic inclusion of native human sera in general clinical chemistry.Frontiers in molecular biosciences · 2026
- A Multi-Laboratory, Multi-Platform Analysis of the Multi-Attribute MethodPharmaceuticals · 2025
- B-039 Harmonizing M-protein analysis: an international, multi-center reproducibility study of the EXENT® System and evaluation of concordance to electrophoretic methodsClinical Chemistry · 2026
- Molecular Attribution and Assay Replacement Readiness of Multi-Attribute Methods in Biosimilar Development.Pharmaceuticals (Basel, Switzerland) · 2026
- Harmonizing M-protein analysis: an international, multi-center reproducibility study of mass spectrometry and evaluation of concordance to electrophoretic methods.Clinical chemistry and laboratory medicine · 2026
- B-064 Bridging Quantitative MALDI-TOF Mass Spectrometry to IMWG Complete Response: Deriving a Serum Cut-off Equivalent to Negative ImmunofixationClinical Chemistry · 2026