Harmonizing M-protein analysis: an international, multi-center reproducibility study of mass spectrometry and evaluation of concordance to electrophoretic methods.
Katie Thoren, Lauren Campbell, Rosa Pello, María Ángeles Iglesias, Raúl Mateos Pablos, Cristina Agulló, Sergio Castro, Shelby M Hutcherson,
Miller School of Medicine, University of Miami, Miami, FL, USA. · Immunology Laboratory, Oxford University Hospitals NHS Foundation Trust, Oxford, UK.
Key points
From the paper's conclusion- Automated MALDI-TOF MS demonstrated strong inter-site reproducibility and close concordance with electrophoretic methods. Residual variability was attributable to upstream immunoglobulin quantification, IgM-related effects and known limitations of electrophoretic integration. These findings support MS as a robust and harmonizable platform for M-protein analysis in multi-center settings.
Abstract
OBJECTIVES: Serum M-protein analysis is central to the assessment of monoclonal gammopathies. Quantitation and isotyping are commonly performed by SPE or CZE and IFE or IT, respectively. While these assays are widely used, they are characterized by limited sensitivity and subjective interpretation. We evaluated the inter-site reproducibility of an automated MALDI-TOF mass spectrometry (MS) platform and its concordance with electrophoretic methods. METHODS: We prepared 120 blinded, serially diluted serum specimens from 17 patients with confirmed monoclonal proteins, of which 118 were eligible for quantitative analyses. Five academic laboratories analyzed identical samples using routine SPE/CZE, IFE/IT, Optilite® total immunoglobulin measurements and MALDI-TOF MS (the EXENT® System). We assessed mass precision, quantitative and qualitative agreement, inter-site reproducibility, and CV-based variability. RESULTS: Mass precision was excellent with >99 % of results within 4 m/z of the five-site mean. Agreement with IFE/IT was 81.7 %. Bland-Altman analysis resulted in mean biases ranging from -1.14 g/L to 4.61 g/L with isotype-dependent differences observed. Global comparison of SPE/CZE to MS resulted in a median Passing-Bablok slope of 1.004 (IQR: 0.055). Inter-site reproducibility was improved for MS over SPE/CZE with median slopes of 1.029 (IQR: 0.083) and 0.997 (IQR: 0.113), respectively. Results with CV ≤15 % were more frequent for MS than by SPE/CZE (67.8% vs. 46.9 %), particularly at MP concentrations <10 g/L. CONCLUSIONS: Automated MALDI-TOF MS demonstrated strong inter-site reproducibility and close concordance with electrophoretic methods. Residual variability was attributable to upstream immunoglobulin quantification, IgM-related effects and known limitations of electrophoretic integration. These findings support MS as a robust and harmonizable platform for M-protein analysis in multi-center settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-039 Harmonizing M-protein analysis: an international, multi-center reproducibility study of the EXENT® System and evaluation of concordance to electrophoretic methodsClinical Chemistry · 2026
- Performance evaluation of an automated mass spectrometry system for the measurement and isotyping of monoclonal immunoglobulinsClinica Chimica Acta · 2026
- Automated peak detection, isotyping and quantitation of M-proteins by MALDI-TOF MS for replacement of immunofixation and serum protein electrophoresisJournal of Immunological Methods · 2026
- B-064 Bridging Quantitative MALDI-TOF Mass Spectrometry to IMWG Complete Response: Deriving a Serum Cut-off Equivalent to Negative ImmunofixationClinical Chemistry · 2026
- B-196 Comprehensive clinical validation of the Immunoglobulin Isotypes (GAM) assay for the EXENT® Analyzer assay in monoclonal gammopathiesClinical Chemistry · 2026
- Mass spectrometry-based screening for monoclonal gammopathy in individuals undergoing routine physical examinations: Analyses of consecutive 15,068 cases in China.Clinica chimica acta; international journal of clinical chemistry · 2026