Clinical Chemistry

B-064 Bridging Quantitative MALDI-TOF Mass Spectrometry to IMWG Complete Response: Deriving a Serum Cut-off Equivalent to Negative Immunofixation

Cristina Agullo Roca, Sergio Castro, Itxaso Jiménez, Tiago Ramalho, Beatriz Rey, Verónica González-Calle, Borja Puertas, Elena Alejo,

Observational · n=87

Key points

From the paper's conclusion
  • We therefore aimed to derive a serum EXENT concentration threshold associated with negative serum IFE, as a potential “IFE-negative equivalent” to support assignment of IMWG CR when IFE is not performed.

Abstract

Background According to the International Myeloma Working Group (IMWG), complete response (CR) requires negative serum and urine immunofixation electrophoresis (IFE), disappearance of any soft-tissue plasmacytomas, and <5% plasma cells in the bone marrow. IFE is a qualitative technique widely used for the detection and monitoring of monoclonal proteins (MP) in monoclonal gammopathies; however, it is labor-intensive and may limit laboratory efficiency. As clinical laboratories consider replacing IFE with automated MALDI-TOF mass spectrometry (MS) workflows, a practical challenge arises: IFE provides a categorical positive/negative result, whereas MS generates quantitative measurements. The EXENT Immunoglobulin Isotypes (GAM) assay is a quantitative MALDI-TOF MS immunoassay designed to identify and semi-quantify serum monoclonal IgG/IgA/IgM in g/L. We therefore aimed to derive a serum EXENT concentration threshold associated with negative serum IFE, as a potential “IFE-negative equivalent” to support assignment of IMWG CR when IFE is not performed. Methods We retrospectively analyzed 87 paired serum samples, each obtained from a different patient, collected as part of routine clinical laboratory practice, with available MP QIP-MS quantification (g/L) using EXENT® system (The Binding Site, part of Thermo Fisher Scientific) and corresponding serum IFE results, obtained using the Vitresi® system (Menarini Diagnostics). IFE results were classified as positive or negative for MP and used as the reference method. QIP-MS concentrations were explored to derive a candidate threshold associated with negative IFE. Based on this exploratory analysis, a QIP-MS cut-off of 0.115 g/L or lower was proposed. Diagnostic performance was assessed by calculating sensitivity, specificity, negative predictive value (NPV), and positive predictive value (PPV). Receiver operating characteristic (ROC) analysis was performed, and 95% confidence intervals (CI) were calculated. Results Of the 87 samples analyzed, 35 were IFE-positive and 52 were IFE-negative. Using the proposed QIP-MS cut-off of 0.115 g/L or lower (AUC: 0.8137; p<0.0001), 34 of 35 IFE-positive samples were correctly identified, yielding a sensitivity of 97% (95% CI: 85.47–99.85) and a NPV of 96.5% (95% CI: 82.82–99.82). The single IFE-positive sample with a QIP-MS concentration below the cut-off (0.09 g/L) corresponded to a weakly detected IgA MP without an associated monoclonal light chain. Specificity was 53.8%, and the PPV was 58.6%. Discordant results were predominantly IFE-negative samples with EXENT concentrations >0.115 g/L, characterized by low-level MS signals close to the threshold. This pattern is consistent with published comparisons showing that MS-based M-protein can be analytically more sensitive than IFE in serum, particularly during treatment monitoring, oligosecretory patterns, immunoglobulin polyclonality, antibody-related IFE nonspecificities, or early clonal activity not detectable by IFE. This pattern explains the observed specificity and PPV values while supporting the role of QIP-MS as a sensitive quantitative tool. Conclusion In this exploratory cohort, we derived an EXENT serum cut-off of 0.115 g/L or lower that was associated with negative serum IFE and demonstrated high sensitivity and NPV, supporting its potential use as a quantitative “serum IFE-negative equivalent” to enable assignment of the laboratory component of IMWG classic CR in MS-only monitoring workflows. The modest specificity reflects the expected impact of higher analytical sensitivity with MS and underscores that full IFE-to-MS replacement may change how “negative” is operationally defined. Larger, independent cohorts (and separate evaluation of the urine component of IMWG CR) are warranted to validate clinical applicability.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers