CVIR Endovascular

Dynamic J-shape steering with a steerable microcatheter: technique and step-specific outcomes in 23 target-branch catheterizations

Kenichiro Okumura, Takahiro Ogi, Junichi Matsumoto, Nobuyuki Asato, Kazuto Kozaka, Satoshi Kobayashi

Observational

Key points

From the paper's conclusion
  • Postprocedural angiography and dynamic contrast-enhanced CT follow-up extending at least 3 months showed no maneuver-related vascular abnormality in any case.

Abstract

Purpose To describe a temporary reverse-J maneuver using a steerable microcatheter and report step-specific outcomes. Materials and methods This retrospective series included 23 target-branch catheterizations performed from March 2020 to March 2026. The reverse-J maneuver was performed in every case, including all four unsuccessful cases; none was completed using direct steering alone. The deflected catheter and guidewire were withdrawn together for branch engagement. Deflection was maintained during guidewire advancement and released after adequate distal wire purchase. Technical success required advancement of the LEONIS Mova tip into the target branch. Results Guidewire entry was achieved in 21/23 cases (91.3%), and LEONIS Mova advancement in 19/23 (82.6%; exact 95% confidence interval, 61.2–95.0). In two cases, the catheter tip could not engage the target ostium and the guidewire could not enter the branch. In two others, the guidewire entered but the catheter could not advance during the transition to the released configuration. No catheter damage, vasospasm, dissection, or difficulty with catheter removal occurred. Postprocedural angiography and dynamic contrast-enhanced CT follow-up extending at least 3 months showed no maneuver-related vascular abnormality in any case. Conclusion Temporary reverse-J formation followed by deflection release enabled catheterization in 19 of 23 attempts.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers