Improving practical steerability of a bidirectional steerable catheter for ERCP by non-straight wire routing.
Atsushi Yamada, Wataru Yonemichi, Ikuo Tooyama, Osamu Inatomi
Medical Innovation Research Center, Shiga University of Medical Science, Otsu, Shiga, 520-2192, Japan. ayamada@belle.shiga-med.ac.jp. · Technology Development Group, Zeon Medical Inc, Takaoka, Toyama, 933-0981, Japan.
Key points
From the paper's conclusion- These findings indicate that non-straight wire-routing architectures can be practically implemented in a bidirectional SC for ERCP and may improve steerability under ERCP-relevant constraints.
Abstract
Endoscopic retrograde cholangiopancreatography (ERCP) requires reliable bile duct cannulation, yet steerability of wire-driven steerable catheters (SCs) can deteriorate when the shaft is constrained, partly because actuation may induce unintended shaft-side deformation and transmission loss. This study evaluated whether non-straight wire-routing architectures could improve practical steerability of a bidirectional SC for ERCP. Three prototypes were developed: a conventional straight-wiring design (SC#0), a non-straight, non-crossed design (SC#1), and a non-straight, crossed design (SC#2). Steerability was assessed by incremental wire-pulling tests under straight-shaft and loose-shaft conditions and by a mock ERCP study using an actual duodenoscope. Outcomes included the handle displacement and pulling force required to reach 85° distal bending (HD85, HPF85) and target-angle reachability. In the straight-shaft condition, SC#1 and SC#2 reduced HD85 and HPF85 by 54-63% and 63-65%, respectively, compared with SC#0. Under the loose-shaft condition, SC#0 reached 85° in only 1/5 trials, whereas SC#1 and SC#2 reached 85° in all trials. In the mock ERCP study, the proposed designs reduced HD85 and HPF85 by 57-71% and 57-62%, respectively, in both bending directions. These findings indicate that non-straight wire-routing architectures can be practically implemented in a bidirectional SC for ERCP and may improve steerability under ERCP-relevant constraints.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multi‐Loop Traction Device‐Assisted Papillary Traction During Single‐Balloon Enteroscopy‐Assisted Endoscopic Retrograde Cholangiopancreatography After Roux‐en‐Y Gastrectomy: A Four‐Case SeriesDEN Open · 2026
- Efficacy and Safety of a Novel Guidewire for Selective Biliary Cannulation: A Multicenter, Prospective Randomized Controlled StudyThe Korean Journal of Pancreas and Biliary Tract · 2026
- Successful Colonic Self-Expandable Metal Stent Placement Using a Rotatable Sphincterotome After Conventional Guidewire Insertion Failure: A Case Report.Cureus · 2026
- Extreme Anticlockwise Endoscopist Repositioning for Biliary Cannulation in Left-wall Periampullary Diverticulum.Endoscopy international open · 2026
- Dynamic J-shape steering with a steerable microcatheter: technique and step-specific outcomes in 23 target-branch catheterizationsCVIR Endovascular · 2026
- Difficult biliary cannulation during ERCP: techniques and an evidence-informed practical algorithm.Frontiers in medicine · 2026