Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions

When the RotaWire Crosses but Cannot Advance: Dual-Lumen Microcatheter-Assisted RotaWire Delivery.

Kazumasa Kurogi, Masanobu Ishii, Nobuyasu Yamamoto, Kenichi Tsujita

Department of Cardiology, Miyazaki Prefectural Nobeoka Hospital, Nobeoka, Japan. · Department of Cardiovascular Medicine, Graduate School of Medical Sciences, Kumamoto University, Kumamoto, Japan.

Case report

Key points

From the paper's conclusion
  • We report a 76-year-old man with a severely calcified mid-left anterior descending coronary artery (LAD) lesion in which neither a microcatheter nor a 0.75 mm balloon could cross.

Abstract

RotaWire delivery can be challenging in severely calcified device-uncrossable coronary lesions when conventional wire exchange through a microcatheter is not feasible. We report a 76-year-old man with a severely calcified mid-left anterior descending coronary artery (LAD) lesion in which neither a microcatheter nor a 0.75 mm balloon could cross. A second channel was created using a penetration wire through a dual-lumen microcatheter, allowing direct RotaWire crossing. However, the RotaWire could not be advanced sufficiently into the distal LAD. After intentional proximal pullback of the dual-lumen microcatheter, the microcatheter and RotaWire were held together and advanced as a single unit, enabling adequate distal RotaWire positioning and successful rotational atherectomy. This case highlights a distinct challenge after successful direct RotaWire crossing and demonstrates coupled advancement with a dual-lumen microcatheter as a potential bailout technique when distal RotaWire advancement remains difficult.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

When the RotaWire Crosses but Cannot Advance: Dual-Lumen Microcatheter-Assisted RotaWire Delivery — Science Skim