The trap device retrieves the detached stent that is floating within the ascending aorta.
Li-An Liao, Meng Liang, Rong-Tai Luo, Hua-Qing Yao, Xin-Ping Lan
Department of Interventional Radiology, Meizhou People's Hospital, Meizhou, Guangdong Province, China. · Meizhou People's Hospital Cardiac Intensive Care Unit, Meizhou, Guangdong Province, China.
Key points
From the paper's conclusion- We reported a patient who previously underwent stent implantation in the left anterior descending artery (LAD) and required a second-stage surgery.
Abstract
BACKGROUND: Coronary stent dislodgement is a relatively infrequent iatrogenic complication associated with percutaneous coronary intervention (PCI). It can result in severe outcomes, including vascular embolism, coronary rupture, malignant arrhythmia, and even death. We reported a patient who previously underwent stent implantation in the left anterior descending artery (LAD) and required a second-stage surgery. During the insertion of a new stent into the left circumflex artery (LCX), the guidewire passed through the mesh of the original stent. This caused the new stent to pull and deform the original stent during advancement. Consequently, the new stent dislodged, and the two stents intertwined and floated in the aorta. Eventually, the two stents were successfully removed using a stent snare, and a new stent was reimplanted in the LAD, thereby avoiding the risk of the dislodged stent falling into other arteries and causing embolism. In our case, the characteristic of the original stent protruding into the left main coronary artery (LM) and the surgeon's failure to promptly confirm the position of the guidewire were the main causes of this incident, which provide valuable experience for future interventional surgeries. DISCUSSION: When the original stent located at the bifurcation of the blood vessel is passed through, preoperative assessment of the spatial relationship between the blood vessels, careful confirmation of the position of the guidewire, and timely analysis of whether stent dislodgement, traction, etc. occurs when advancement is impeded are critical to procedural success.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Double coronary stent loss with aortic migration and iliac embolization successfully managed by snare retrieval: a case report.Frontiers in cardiovascular medicine · 2026
- Successful Bailout of Guidewire Entrapment Within a Coronary Stent Using a Tornus Penetration Catheter: A Case Report.Cureus · 2026
- Morphology Matters: Persistent Iatrogenic Aorto-Coronary Dissection Despite Initial Sealing Treated with a Stent-in-Stent Bailout Strategy: A Case Report and Literature Review.Reports (MDPI) · 2026
- When Subtle Guiding Catheter-Induced Stent Deformation Leads to Abluminal Guidewire Passage: Successful IVUS Tip-Detection-Guided Bailout.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Coronary stent fixation of left ventricular lead during CRT-D implantation: a case series of three patients.Frontiers in cardiovascular medicine · 2026
- Balloon-Stopper Retrieval and Peripheral Parking After Inadvertent Ostial Right Coronary Stent Deployment.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026