Balloon-Stopper Retrieval and Peripheral Parking After Inadvertent Ostial Right Coronary Stent Deployment.
Fernando Quevedo-Candela, Guy Proulx, Tomás Cieza
Interventional Cardiology Division, Quebec Heart-Lung Institute, Laval University, Québec, Canada.
Key points
From the paper's conclusion- We describe a 70-year-old man undergoing PCI for an ostial RCA lesion where a forceful cough caused complete stent expansion within the aortic root.
Abstract
BACKGROUND: Percutaneous treatment of right coronary artery (RCA) ostial lesions is technically challenging due to cardiorespiratory motion. CASE SUMMARY: We describe a 70-year-old man undergoing PCI for an ostial RCA lesion where a forceful cough caused complete stent expansion within the aortic root. The stent remained mechanically engaged to the delivery shaft. A 4.0 × 15 mm balloon was inflated at the guiding catheter tip as a "balloon stopper," enabling controlled retrieval and "peripheral parking" (intentional implantation) in the right brachial artery. DISCUSSION: This case illustrates the technical feasibility of a bailout strategy for selected fully expanded extracoronary stents when guidewire access is maintained. The balloon-stopper technique provided mechanical control during retrieval, while peripheral implantation allowed relocation of the stent when further extraction was considered potentially hazardous.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Percutaneous Coronary Stentectomy: Redefining Stent Failure Management in the Modern Era.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Double coronary stent loss with aortic migration and iliac embolization successfully managed by snare retrieval: a case report.Frontiers in cardiovascular medicine · 2026
- Guide Extension and Balloon-Assisted Retrieval of an Entrapped Embolic Protection Device Within a Coronary Stent: A Case Report and In Vitro Analysis.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Percutaneous Retrieval of a Hepatic Vein Stent From the Right Ventricle: The triPle-rAil snare Technique.JACC. Case reports · 2026
- When an 8-mm Coronary Artery Has No Appropriate Stent: Bailout PCI and Dual-Balloon Optimization in Giant Right Coronary Artery Ectasia.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Bailout Retrieval of a Ruptured Entrapped Coronary Balloon Using the Reinflation-Deflation Technique (Nantapon-Wittawat Technique).Cureus · 2026