Successful Bailout of Guidewire Entrapment Within a Coronary Stent Using a Tornus Penetration Catheter: A Case Report.
Tomohide Sato, Eiji Horio, Kazuhisa Kodama, Tomohiro Sakamoto, Junjiroh Koyama
Cardiology, Saiseikai Kumamoto Hospital Cardiovascular Center, Kumamoto, JPN.
Key points
From the paper's conclusion- This case demonstrates a potential bailout technique for guidewire entrapment within a coronary stent.
Abstract
Guidewire entrapment within an implanted coronary stent is a rare but potentially serious complication of percutaneous coronary intervention (PCI), and its management can be technically challenging. An 81-year-old woman presenting with acute myocardial infarction underwent PCI for a severely stenotic lesion in the right coronary artery. After drug-eluting stent implantation and post-dilatation, intravascular ultrasound (IVUS) revealed deformation at the distal edge of the stent. During withdrawal of the IVUS catheter, the catheter and guidewire became entangled with the deformed stent. Although the IVUS catheter was successfully removed with careful manipulation, the guidewire remained firmly entrapped within the stent and could not be retrieved despite repeated attempts. A Tornus penetration catheter was subsequently advanced over the entrapped guidewire using rotational manipulation toward the site of entrapment. When the catheter reached the entrapped segment, the guidewire was successfully disengaged from the stent and retrieved. Following successful bailout, distal stent fracture and proximal coronary dissection were identified, requiring additional stent implantation. This case demonstrates a potential bailout technique for guidewire entrapment within a coronary stent. Rotational advancement of a penetration catheter to the site of entrapment may facilitate disengagement of an entrapped guidewire when standard manipulation is unsuccessful.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Guide Extension and Balloon-Assisted Retrieval of an Entrapped Embolic Protection Device Within a Coronary Stent: A Case Report and In Vitro Analysis.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Reverse Balloon Anchoring for Retrieval of an Entrapped Side-Branch Balloon During the Jailed Balloon Technique.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Single-Access Retrieval of an Entrapped Rotational Atherectomy Burr Using Parallel-Wire STAR Technique.JACC. Case reports · 2026
- Endovascular Retrieval of Unraveled Coils Extending Along the Aorta Due to Guidewire Fracture During Percutaneous Coronary Intervention: The Snare and Trap Technique.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Wire-entanglement retrieval of a detached rotational atherectomy burr after complete wire disengagement: a case reportEuropean Heart Journal - Case Reports · 2026
- When Subtle Guiding Catheter-Induced Stent Deformation Leads to Abluminal Guidewire Passage: Successful IVUS Tip-Detection-Guided Bailout.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026