Coronary stent fixation of left ventricular lead during CRT-D implantation: a case series of three patients.
Hongwei Yi, Jia Li, Yue Bao, Hongwei Han
Department of Cardiology, Wuhan Asian Heart Hospital, Wuhan, Hubei, China.
Key points
From the paper's conclusion- Objective To investigate a salvage method using coronary stent fixation of the left ventricular (LV) lead to manage refractory LV lead dislodgement caused by unfavorable coronary venous anatomy during cardiac resynchronization therapy defibrillator (CRT-D) implantation, and to evaluate its technical feasibility and clinical value.
Abstract
OBJECTIVE: To investigate a salvage method using coronary stent fixation of the left ventricular (LV) lead to manage refractory LV lead dislodgement caused by unfavorable coronary venous anatomy during cardiac resynchronization therapy defibrillator (CRT-D) implantation, and to evaluate its technical feasibility and clinical value. METHODS: We retrospectively included three patients with heart failure with reduced ejection fraction (HFrEF) at our center who developed LV lead dislodgement during or early after CRT-D implantation. Conventional methods failed to maintain a stable lead position. A sirolimus-eluting coronary stent was deployed within the target coronary vein to achieve mechanical anchoring of the LV lead via gentle compression against the venous wall. Procedural details, perioperative complications, and follow-up data were collected and analyzed. RESULTS: All three patients completed stent fixation of the LV lead. No acute severe complications, such as coronary venous rupture, pericardial tamponade, or lead conductor damage, occurred intraoperatively. No significant change in pacing threshold was observed before and after stent deployment. Cases 1 and 2 were followed for 30 and 26 months, respectively, and lead parameters remained stable. Case 3 was followed for only 2 months and then lost to follow-up; lead parameters were stable at that time point. No clinically meaningful improvement in LV ejection fraction (LVEF) was observed in Cases 1 and 2 during follow-up. Because Case 3 was followed for only 2 months, shorter than the usual 6-12-month window for assessing CRT reverse remodeling, functional response could not be evaluated. This procedure constitutes off-label device use; after full stent deployment, the lead position cannot be readjusted. If device-related infection occurs, transvenous lead extraction may be difficult and uncertain, and surgical removal may be required. CONCLUSIONS: After all conventional methods have failed, stent fixation may serve as a salvage option for refractory LV lead dislodgement caused by unfavorable coronary venous anatomy. This report describes the clinical observations of sirolimus-eluting stent use in this off-label salvage procedure and details the relevant procedural parameters and techniques. However, the sample size is small, and potential long-term risks, including stent fracture, lead insulation damage, and stability of lead electrical parameters, require further evaluation in larger studies. This technique only addresses the mechanical fixation of the lead and does not guarantee functional benefit from CRT.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Bailout Strategy for CRT Implantation: Retrograde Coronary Venous Wiring with Venoplasty and Wire Externalization.Indian pacing and electrophysiology journal · 2026
- Early Right Ventricular Implantable Cardioverter-Defibrillator Lead Microdislodgement Associated With Active-Fixation Helix Retraction: A Case Report.Cureus · 2026
- Double coronary stent loss with aortic migration and iliac embolization successfully managed by snare retrieval: a case report.Frontiers in cardiovascular medicine · 2026
- Novel Use of Intravascular Lithotripsy for the Mobilization of an Entrapped Coronary Guidewire in a Heavily Calcified LAD/DB Bifurcation Lesion-A Case Report.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- The trap device retrieves the detached stent that is floating within the ascending aorta.Journal of cardiothoracic surgery · 2026
- Crossing an Angulated SVG-RCA Anastomosis Using Distal Anchoring and Guide Extension: A Bailout Strategy in STEMI.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026