A biomechanical comparison of closed-reduction osteosynthesis techniques for fractures of the ulnar styloid.
Jonas Eck, Inska Onken, Julian P Maier, Kilian Reising, Nikos Karvouniaris, Jörg Bayer, Hagen Schmal, Ferdinand C Wagner
Department of Orthopedics and Trauma Surgery, University Medical Center Freiburg, Freiburg, Germany. jonas.eck@uniklinik-freiburg.de. · Department of Orthopedics and Trauma Surgery, University Medical Center Freiburg, Freiburg, Germany.
Key points
From the paper's conclusion- Ulnar styloid refixation with a headless compression screw provides superior axial stability compared to single or double K-wire fixation. This should be considered in the treatment of unstable USF.
Abstract
BACKGROUND: Ulnar styloid fractures (USF) are frequently underestimated and can contribute to clinically relevant wrist instability and impaired functional outcomes. Several osteosynthesis techniques are available for the treatment of unstable USF. This study aims to compare the biomechanical stability of fixation methods applicable to closed reduction - specifically K-wire and headless compression screw fixation - to support clinical decision-making. METHODS: This biomechanical in vitro study was conducted using 30 specimens of artificial ulnae. USF were simulated through a standardized osteotomy at the base. Three groups of ten specimens each were created to compare three methods of osteosynthesis: One 1.4 mm (mm) Kirschner-wire (K-wire) (Group 1KW), two 1.0 K-wires (Group 2KW) or one 2.0 cannulated headless screw (Group CHS). Each construct underwent 1,000 load cycles up to 50 newtons (N) followed by load-to-failure testing, which was defined as displacement ≥ 2 mm. Displacement [mm], stiffness [N/mm] and load to failure [N] were recorded and compared among the groups. RESULTS: The mean fracture displacement after 1,000 cycles was 0.58 mm (SD 0.37) in the 1KW group, 0.23 mm (SD 0.09) in the 2KW group and 0.09 mm (SD 0.03) in the CHS group. Significant group differences could be shown between CHS and 1KW (p < 0.01) as well as between CHS and 2KW (p = 0.017). The mean construct stiffness was 269.95 (SD 45.72) N/mm in the 1KW group, 484.08 (SD 207.60) N/mm in the 2KW group and 730.95 (SD 241.42) N/mm in the CHS group. Significant group differences could be shown between CHS and 1KW (p < 0.01). The mean load-to-failure was 136.82 (SD 68.86) N in the 1KW group, 181.60 (SD 28.09) N in the 2KW group and 204.17 (SD 63.10) N in the CHS group. Significant group differences could be shown between CHS and 1KW (p = 0.013). CONCLUSION: Ulnar styloid refixation with a headless compression screw provides superior axial stability compared to single or double K-wire fixation. This should be considered in the treatment of unstable USF.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Clinical Outcomes of Single Intramedullary K-Wire Fixation with Figure-of-Eight Suture Augmentation for Ulnar Styloid Base and Distal Ulnar Metaphyseal Fractures.Journal of clinical medicine · 2026
- Biomechanical comparison of palmar plate plus headless compression screw versus radiopalmar double plating in AO/OTA 23-C2.1 distal radius fracturesArchives of Orthopaedic and Trauma Surgery · 2026
- Biomechanical comparison between two headless screw configurations for fixation of coronal shear capitellum humeri fractures.Clinical biomechanics (Bristol, Avon) · 2026
- Biomechanical comparison of modified tension band wiring using Kirschner wires with eyelets and conventional tension band wiring for olecranon fractures: A composite bone study.Clinical biomechanics (Bristol, Avon) · 2026
- Biomechanical efficacy of wire fixation methods for oblique metacarpal shaft fracturesMedicine · 2026
- Mini-Anchor Fixation vs Conservative Treatment for Hauck Type II Ulnar Styloid Fractures Associated With Distal Radius Fractures: A Retrospective Cohort Study.Medical science monitor : international medical journal of experimental and clinical research · 2026