Biomechanical comparison of modified tension band wiring using Kirschner wires with eyelets and conventional tension band wiring for olecranon fractures: A composite bone study.

Jung Eun Lee, Do Yeon Kim, Deuk Soo Jun, Young Ho Lee

Department of Orthopedic Surgery, Gachon University Gil Medical Center, Incheon 21565, Republic of Korea. · Department of Orthopedic Surgery, Seoul National University College of Medicine, Seoul National University Hospital, Seoul 03080, Republic of Korea. Electronic address: orthoyhl@snu.ac.kr.

Key points

From the paper's conclusion
  • Under the tested conditions, no statistically significant differences in measured biomechanical outcomes were detected between ring pin TBW and conventional TBW. These findings provide a biomechanical basis for further evaluation of ring pin TBW as an alternative fixation technique, potentially offering technical advantages through intramedullary insertion without anterior cortical penetration.

Abstract

BACKGROUND: Tension band wiring (TBW) is widely accepted for treating simple transverse olecranon fractures but is associated with complications, including wire migration and neurovascular injury. Ring pins with eyelets were developed to address these limitations; nonetheless, biomechanical comparisons with conventional fixation remain limited. We compared both techniques for transverse olecranon fractures in a composite bone model. METHODS: Twenty-two fourth-generation composite ulnae received conventional fixation with Kirschner wires or modified fixation with ring pins (11 per group). Cyclic loading comprised 100 cycles at 10-20 N followed by 1000 cycles at 20-220 N, both at 1 Hz. Testing-machine crosshead displacement was recorded, and regional interfragmentary gap displacement was measured after unloading using a noncontact coordinate-measuring machine. Load-to-failure testing determined maximum failure load. FINDINGS: No statistically significant differences were detected in crosshead displacement after cyclic loading (conventional: 6.29 ± 1.94 versus ring pin: 6.22 ± 1.83 mm; p = 0.898) or maximum failure load (276.91 ± 27.54 versus 276.86 ± 19.83 N; p = 0.898). Mean differences (ring pin minus conventional) were - 0.07 mm (95% confidence interval - 1.75 to 1.61) and - 0.05 N (-21.53 to 21.43), respectively. Regional gap displacement also showed no statistically significant differences (all p > 0.05). INTERPRETATION: Under the tested conditions, no statistically significant differences in measured biomechanical outcomes were detected between ring pin TBW and conventional TBW. These findings provide a biomechanical basis for further evaluation of ring pin TBW as an alternative fixation technique, potentially offering technical advantages through intramedullary insertion without anterior cortical penetration.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers