Reliability of shear wave elastography measurements of the linea alba: an intraclass correlation and Bland-Altman analysis.
Sara Cañamero de León, Mercedes Soto-González, Iria Da Cuña Carrera, Augusto Gil Pascoal
Faculty of Physical Therapy. Department of Functional Biology and Health Science, University of Vigo, Pontevedra, Spain. · Faculty of Physical Therapy. Department of Functional Biology and Health Science, University of Vigo, Pontevedra, Spain. iriadc@uvigo.es.
Key points
From the paper's conclusion- Shear-wave elastography (SWE) demonstrated acceptable-to-good reproducibility under standardized conditions, although reliability varied according to examiner and anatomical location. Averaging three repeated measurements improved intra-rater reliability, while inter-rater reliability was highest at BL2 and lowest at AB5. These findings require confirmation in larger studies.
Abstract
OBJECTIVE: Diastasis recti abdominis (DRA) is a common postpartum condition characterized by an increased separation between the rectus abdominis muscles along the linea alba, which may affect the stability of the abdomino-lumbo-pelvic complex and quality of life. Two-dimensional ultrasound is an established technique for assessing inter-rectus distance, whereas, more recently, shear wave elastography (SWE) has emerged as a promising method for quantifying tissue stiffness, although it is not yet fully established in clinical practice. However, before its routine use, it is necessary to establish the reliability of SWE measurements, considering factors such as anatomical location and examiner experience. This study aimed to determine intra- and inter-rater reliability of shear wave elastography for assessing the linea alba stiffness above and below the umbilicus in women with and without diastasis rectus abdominis. METHODS: An observational, cross-sectional study was conducted. Twenty-nine participants were included and two licensed examiners with different levels of elastography experience performed independent assessments on the same day, acquiring three SWE measurements above (AB5), at (UBO), and below (BL2) the umbilicus using a standardized protocol. Reliability was assessed using ICCs and Bland-Altman analysis. RESULTS: Intra-rater reliability was higher when three repeated measurements were averaged. For Examiner 1, mean-of-three ICCs ranged from 0.839 (95% CI 0.703-0.919) to 0.914 (95% CI 0.842-0.957), whereas for Examiner 2 they ranged from 0.662 (95% CI 0.371-0.831) to 0.800 (95% CI 0.630-0.899). Inter-rater reliability varied by anatomical location. For single examiner-specific mean values, ICC(A,1) was 0.439 (95% CI 0.086-0.692) at AB5, 0.517 (95% CI 0.187-0.741) at UBO, and 0.601 (95% CI 0.313-0.789) at BL2. For the mean of both examiners, ICC(A,2) was 0.610 (95% CI 0.159-0.818), 0.681 (95% CI 0.315-0.851), and 0.751 (95% CI 0.476-0.882), respectively. Bland-Altman analysis indicated minimal systematic bias and acceptable agreement: AB5, + 0.07 m/s (95% LoA - 2.74 to + 2.88); UBO, + 0.09 m/s (95% LoA - 2.63 to + 2.82); and BL2, - 0.21 m/s (95% LoA - 2.04 to + 1.63). Mean absolute errors ranged from 0.70 to 1.13 m/s, with no evidence of proportional bias. CONCLUSION: Shear-wave elastography (SWE) demonstrated acceptable-to-good reproducibility under standardized conditions, although reliability varied according to examiner and anatomical location. Averaging three repeated measurements improved intra-rater reliability, while inter-rater reliability was highest at BL2 and lowest at AB5. These findings require confirmation in larger studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effects of operator experience and measurement depth on skeletal muscle shear wave elastography.Journal of ultrasound · 2026
- Intra- and Inter-Rater Reliability of Inter-Recti Distance Measurement Using B-Mode and Panoramic Ultrasound ImagingJournal of Clinical Medicine · 2026
- Measurement Reliability and Trial-Sampling Effects in Ultrasound Assessment of Abdominal Muscle Thickness During the Active Straight Leg RaiseDiagnostics · 2026
- Reproducibility of Cervical Strain Elastography Using a Standardized Acquisition Protocol in Mid-Trimester Pregnancy.Journal of clinical ultrasound : JCU · 2026
- Assessment of placental stiffness during labor using ultrasound shear-wave elastography: a pilot study.Placenta · 2026
- Application of three-dimensional visualization reconstruction technology through CT to quantitatively measure diastasis recti abdominisBMC Medical Imaging · 2026