Effect of metal sleeve integration on implant placement accuracy using additively manufactured tooth-supported surgical guides: an in vitro study
Dariely Ramos Guzmán, Fernando Luengo Mas, Pablo Sevilla Hernández, Elisabeth Casañas Gil, María Vinyet Ris Casado, Ignacio García-Gil, Pablo Romero Villalba, Jaime Jiménez García
Key points
From the paper's conclusion- Background This in vitro study evaluated whether metal sleeve integration influences crestal horizontal, crestal vertical, and angular implant-placement deviations when additively manufactured tooth-supported surgical guides are used.
Abstract
Background This in vitro study evaluated whether metal sleeve integration influences crestal horizontal, crestal vertical, and angular implant-placement deviations when additively manufactured tooth-supported surgical guides are used. Methods Thirty-two partially edentulous maxillary resin models were allocated to two groups: tooth-supported surgical guides with metal sleeves and sleeveless guides ( n = 16 per group). Both guide designs were generated from the same virtual implant plan and manufactured using LCD-based vat photopolymerization. A VEGA implant measuring 4.0 × 10 mm was placed at the maxillary right first molar (FDI 16) in each model under standardized laboratory conditions. After implant placement, scan abutments were connected and the models were digitized using a laboratory scanner. The postoperative STL files were aligned with a master reference STL generated from a physical reference model intended to reproduce the planned implant position. Crestal horizontal, crestal vertical, and angular deviations relative to this master reference were measured. Between-group differences were assessed using the Mann–Whitney U test, with the significance level set at α = 0.05. Results No statistically significant differences were identified between the sleeveless and metal-sleeved groups for crestal horizontal deviation (median: 0.36 vs. 0.28 mm; p = 0.287), crestal vertical deviation (median: 0.53 vs. 0.63 mm; p = 0.402), or angular deviation (median: 8.00° vs. 10.00°; p = 0.171). Conclusions Under the conditions of this in vitro study, no statistically significant differences were detected between the metal-sleeved and sleeveless groups in crestal horizontal, crestal vertical, or angular deviation. Because the study was designed to test between-group differences rather than equivalence, these findings should not be interpreted as evidence that the two guide designs are equivalent. Further studies using prespecified equivalence margins and clinical settings are required.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy of Metallic-Sleeved Versus Sleeve-Free Surgical Guides in Static Computer-Assisted Implant Surgery: A Systematic Review and Meta-Analysis.Journal of clinical medicine · 2026
- Does manufacturing specialization matter? Transfer accuracy of static surgical guides produced by general-purpose and dental-specific additive workflows: A paired in vitro study.Journal of clinical and experimental dentistry · 2026
- Comparative accuracy of dynamic navigation versus static surgical guides for dental implant placement: an in vitro assessment of free-end and non-free-end configurationsScientific Reports · 2026
- Impact of Level of Guidance on the Accuracy of Static Computer-Assisted Implant Surgery (s-CAIS): An In Vitro Study.Clinical oral implants research · 2026
- Accuracy of an innovative extraction socket-supported guide for immediate anterior implantation: An in vitro study.Journal of prosthodontics : official journal of the American College of Prosthodontists · 2026
- Accuracy of digital surgical guides for dental implant placement: A systematic review, meta-analysis and region-specific analysis of the anterior maxilla.International journal of oral implantology (Berlin, Germany) · 2026