Comparative accuracy of dynamic navigation versus static surgical guides for dental implant placement: an in vitro assessment of free-end and non-free-end configurations
Chang De Sui, Fatanah Mohamad Suhaimi, Juzailah Roffie, Ahmad Fakrurrozi Mohamad
Key points
From the paper's conclusion- Within the limits of this single-material in vitro model, dynamic navigation was more accurate than static surgical guides, with the largest advantage (40–46% improvement) in free-end sites where guide support is compromised.
Abstract
To compare the accuracy of dynamic navigation and static surgical guides for dental implant placement, with emphasis on free-end versus non-free-end configurations. An in vitro study was conducted using stereolithographic models of partially edentulous maxillary and mandibular arches. Sixty implant sites were virtually planned in 3Shape software and allocated to static surgical guides ( n = 30) or dynamic navigation ( n = 30), each stratified by support configuration (free-end vs. non-free-end) and jaw location (maxilla vs. mandible). Dentium Superline implants (4.0 × 10.0 mm) were placed using 3D-printed static guides or a dynamic navigation system. CBCT scans were acquired before and after placement, and angular (°), entry-point (mm) and apical (mm) deviations were calculated. Dynamic navigation was significantly more accurate than static guides overall (angular 1.95°±0.24° vs. 2.68°±0.54°; entry-point 1.51 ± 0.18 vs. 2.24 ± 0.51 mm; apical 1.75 ± 0.21 vs. 2.51 ± 0.52 mm; all p < 0.001). Pooled across modalities, free-end sites showed higher deviations than non-free-end sites for all three metrics ( p < 0.01); this difference was confined to static guides (angular 3.16° vs. 2.19°) and was not seen with dynamic navigation (1.89° vs. 2.00°). In free-end sites, dynamic navigation reduced angular, entry-point and apical deviations by 40.2% (1.89°±0.23° vs. 3.16°±0.24°), 46.1% (1.46 ± 0.17 vs. 2.71 ± 0.20 mm) and 43.0% (1.70 ± 0.20 vs. 2.98 ± 0.22 mm), respectively (all p < 0.001). In non-free-end sites the advantage was smaller (8.7–12.4%) but significant ( p = 0.032, 0.002 and 0.005). A significant guidance modality × support configuration interaction was found for all outcomes ( p < 0.001; partial η² 0.59–0.68). Jaw location had no significant effect ( p > 0.05). Within the limits of this single-material in vitro model, dynamic navigation was more accurate than static surgical guides, with the largest advantage (40–46% improvement) in free-end sites where guide support is compromised. Dynamic navigation may therefore be preferable for distal extension cases, pending clinical validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Prospective analysis of site-specific accuracy and error sources in dynamically navigated implant surgery.Journal of prosthodontics : official journal of the American College of Prosthodontists · 2026
- Accuracy of Planned Versus Actual Implant Placement Using Different Implant Systems: A Systematic Review of Angulation and Positional Deviations.Journal of maxillofacial and oral surgery · 2026
- Three-Dimensional Accuracy of Static Guides and Dynamic Navigation for Palatal Miniscrew Insertion: A Systematic Review of Template Fabrication and Clinical WorkflowsBioengineering · 2026
- Impact of Level of Guidance on the Accuracy of Static Computer-Assisted Implant Surgery (s-CAIS): An In Vitro Study.Clinical oral implants research · 2026
- Accuracy of Computer-Guided Dental Implant Placement: A Clinical Comparison of Three Surgical Guide TypesJournal of Clinical Medicine · 2025
- Does manufacturing specialization matter? Transfer accuracy of static surgical guides produced by general-purpose and dental-specific additive workflows: A paired in vitro study.Journal of clinical and experimental dentistry · 2026