Journal of prosthodontics : official journal of the American College of Prosthodontists

Accuracy of an innovative extraction socket-supported guide for immediate anterior implantation: An in vitro study.

Jianye Song, Zuge Yang, Wenhao Yang, Zhonghua Cao, Zhenxin Huang, Fangke Jiang, Weiwei Guo, Kun Wang

Department of Stomatology, Xinqiao Hospital, Army Medical University, Chongqing, P. R. China. · State Key Laboratory of Oral & Maxillofacial Reconstruction and Regeneration, National Clinical Research Center for Oral Diseases, Shaanxi Clinical Research Center for Oral Diseases, Department of Oral and Maxillofacial Surgery, School of Stomatology, The Fourth Military Medical University, Xi'an, Shaanxi, P. R. China.

Preclinical

Key points

From the paper's conclusion
  • ESSGs demonstrated greater precision than TSSGs in resin models simulating labial and palatal deviations of socket morphology. Nevertheless, clinical studies are still needed to evaluate their applicability in patients.

Abstract

PURPOSE: This study aimed to evaluate the accuracy of a root-shaped extraction socket-supported guide (ESSG) for anterior single-tooth immediate implant placement. MATERIALS AND METHODS: After the central incisors were digitally removed from cone beam computed tomography (CBCT) images obtained from patients, 3D-printed models were created to simulate labial (Type I), central (Type II), and palatal root-position deviation (Type III) in the sagittal plane. Resin models of each type were divided into high-elastic-modulus (HE) and normal-elastic-modulus (NE) subgroups. In total, 120 implants were immediately placed using ESSGs (test group) or tooth-supported surgical guides (TSSGs; control group). The angular and positional deviations were assessed by comparing post-placement CBCT scans with the corresponding virtual implant plans. RESULTS: In Type I and III resin models, angular deviations were significantly smaller in the test group (p < 0.05) than in the control group, with values of 1.34° ± 0.75° versus 4.28° ± 0.71° (Type I, HE), 0.83° ± 0.76° versus 2.74° ± 1.05° (Type I, NE), 1.18° ± 0.82° versus 3.57° ± 1.40° (Type III, HE), and 1.17° ± 0.79° versus 3.41° ± 1.16° (Type III, NE), respectively. Positional deviations in most Type I and III resin models were likewise significantly lower in the test group than in the control group. Notably, ESSGs maintained high accuracy in HE resin models of Types I and III. CONCLUSIONS: ESSGs demonstrated greater precision than TSSGs in resin models simulating labial and palatal deviations of socket morphology. Nevertheless, clinical studies are still needed to evaluate their applicability in patients.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Accuracy of an innovative extraction socket-supported guide for immediate anterior implantation: An in vitro study — Science Skim