Effects of bone-quality configuration, effective peri-implant interface stiffness, and loading configuration on stress distribution in an ultrashort implant system: a three-dimensional finite element analysis.
Betül Sümeyra Akça, Sercan Küçükkurt
Private Practice in Oral and Maxillofacial Surgery, Istanbul, Türkiye. bsakca55@gmail.com. · Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Istanbul Aydın University, Istanbul, Türkiye.
Key points
From the paper's conclusion- Loading configuration affected component stresses more than nominal effective interface-stiffness scaling. Within this model, increasing nominal effective interface stiffness reduced implant stress, particularly in lower-support configurations, but component stresses remained higher under the oblique-loading configuration. These comparative trends derive from static, linear-elastic, tied-contact simulations without experimental validation and do not define clinical loading protocols, healing periods, survival probabilities, or failure thresholds.
Abstract
BACKGROUND: Ultrashort implants may reduce vertical augmentation needs when residual bone height is limited, but reduced intraosseous support may increase sensitivity to eccentric loading. This full-factorial study evaluated the effects of bone-quality configuration, effective peri-implant interface stiffness, and loading configuration in a restored ultrashort implant. METHODS: A three-dimensional linear static finite element model included a 4.1 × 4.0-mm tissue-level implant, separate abutment and connection screw, and a 7-mm-high molar crown. Four bone-quality configurations (D1-D4), four nominal effective interface-stiffness levels (25%, 50%, 75%, and 100% of adjacent host-bone modulus), and two static 100-N loading configurations-axial (0°) and oblique (30°)-yielded 32 simulations. Stiffness levels were computational surrogates produced by scaling the peri-implant transition-region modulus with tied contacts, not histological osseointegration percentages or healing stages. Component von Mises stress and bone maximum and minimum principal stresses were evaluated. The D1 model at 25% stiffness under the oblique-loading configuration underwent mesh-sensitivity testing at 0.4, 0.3, 0.2, and 0.1 mm. RESULTS: At 100% nominal effective interface stiffness, maximum component von Mises stresses under the oblique-loading configuration were 6.9-9.4 times the axial values. Increasing nominal effective interface stiffness from 25% to 100% reduced maximum implant stress by 4.9%-36.0% and 17.2%-42.0% under the axial- and oblique-loading configurations, respectively, with the largest reduction in D4; abutment and crown stresses changed minimally. In D2-D4, increasing nominal effective interface stiffness decreased cortical maximum principal stress and the magnitude of trabecular minimum principal stress but increased the magnitude of cortical minimum principal stress, indicating redistribution. Final refinement changed implant von Mises and cortical maximum principal stresses by 2.31% and 2.75%, below the 3% threshold. CONCLUSIONS: Loading configuration affected component stresses more than nominal effective interface-stiffness scaling. Within this model, increasing nominal effective interface stiffness reduced implant stress, particularly in lower-support configurations, but component stresses remained higher under the oblique-loading configuration. These comparative trends derive from static, linear-elastic, tied-contact simulations without experimental validation and do not define clinical loading protocols, healing periods, survival probabilities, or failure thresholds.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Initial biomechanical responses during maxillary first molar intrusion across different root-sinus configurations and sinus floor cortical thicknesses: a three-dimensional finite element study.Odontology · 2026
- Comparison of subperiosteal, short, and standard endosseous implants in the severely atrophic edentulous mandible: a three-dimensional finite element stress analysis study.BMC oral health · 2026
- Finite Element Analysis of Ultra-Short, Extra-Short, and Conventional Dental Implants in Fibula Free Flap Mandibular Reconstruction.Bioengineering (Basel, Switzerland) · 2026
- Influence of implant configuration on stress distribution in quad zygoma implant systems: A finite element analysis.Journal of prosthodontics : official journal of the American College of Prosthodontists · 2026
- Influence of implant depth on talus-implant biomechanical behavior in partial talus replacement: a finite element analysis.BMC musculoskeletal disorders · 2026
- Single Versus Double Implants in a Narrowed Mandibular Molar Space: A Finite Element Analysis.Journal of functional biomaterials · 2026