Influence of implant depth on talus-implant biomechanical behavior in partial talus replacement: a finite element analysis.
Yingdong Zhang, Baozhou Zhang, Heng Li, Zeyu Zhang, Xuewen Wang, Yong Wu
Department of Foot and Ankle Surgery, Beijing Jishuitan Hospital, Capital Medical University, Beijing, 102208, China. · Department of Foot and Ankle Surgery, Beijing Jishuitan Hospital, Capital Medical University, Beijing, 102208, China. yongwu11jst@163.com.
Key points
From the paper's conclusion- Implantation depth significantly affected stress and strain in the talus, cartilage, and prosthesis. Recessing the implant by 0.5-1.0 mm below the talar cartilage surface produced a more favorable stress distribution in this single patient-specific geometric model. However, these in silico findings require experimental validation.
Abstract
BACKGROUND: Large, uncontained osteochondral lesions of the talar shoulder (OLTs) remain a surgical challenge. Three-dimensional (3D)-printed partial talar prostheses represent a potential treatment option because, compared with conventional autologous bone grafting, they can restore a smooth articular surface. However, the optimal position of the implant between the cartilage surface and the underlying bone surface remains controversial. Clinical experience suggests that the implant should be placed approximately 1 mm below the cartilage surface, but this empirical recommendation requires quantitative support. METHODS: A 3D finite element analysis (FEA) model comprising the tibia, fibula, talus, calcaneus, talar prosthesis, corresponding cartilage, and ligaments was developed. Cartilage thickness was set to 1.5 mm, and the prosthesis was positioned flush with the cartilage surface (0 mm) or recessed by 0.5, 1.0, or 1.5 mm. The proximal tibia and fibula were constrained, and loading conditions representing bipedal standing and three positions in the gait cycle were applied. For each loading condition, von Mises stress and strain were calculated. RESULTS: Across the four loading conditions, implants recessed by 0.5 and 1.0 mm exhibited more favorable biomechanical behavior. During bipedal standing, maximum von Mises stress at the bone-implant interface was 0.596 and 0.617 MPa for the 0.5- and 1.0-mm configurations, respectively. These configurations also produced more uniform stress distributions during the gait cycle and lower strain magnitudes, suggesting potentially better long-term stability. CONCLUSION: Implantation depth significantly affected stress and strain in the talus, cartilage, and prosthesis. Recessing the implant by 0.5-1.0 mm below the talar cartilage surface produced a more favorable stress distribution in this single patient-specific geometric model. However, these in silico findings require experimental validation. LEVEL OF EVIDENCE: Biomechanical study, Level V. TRIAL REGISTRATION: Not applicable.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Design and biomechanical investigation of a patient-specific total talar prosthesis for total talar replacement: A finite element analysis.The International journal of artificial organs · 2026
- Effects of Talar Osteochondral Lesion Location and Size on the Ankle Joint: A Finite Element Analysis With Cadaveric Validation.Foot & ankle specialist · 2026
- Asymmetric Metaphyseal Cones for AORI Type 2 Medial Defects in Tibial Plateau Fractures during Acute TKAJournal of Orthopaedic Research® · 2026
- A novel strategy for reconstruction of AORI type III proximal tibial defects : surface-type knee reconstruction following defect reconstruction with 3D-printed extension-stem augment.Bone & joint research · 2026
- Effects of bone-quality configuration, effective peri-implant interface stiffness, and loading configuration on stress distribution in an ultrashort implant system: a three-dimensional finite element analysis.BMC oral health · 2026
- Posterior Plate Positioning Without Distal Placement in Medial Closing-Wedge Distal Femoral Osteotomy.Orthopaedic journal of sports medicine · 2026