Comparison of rapid portable devices and laboratory methods to assess the blood chemistry of farmed adult common snook (Centropomus undecimalis).
Cristina Pascual, Pedro Gallardo, Jaime Suárez, Rosario Martínez-Yañez, Ángel Rojo-Cebreros, Honorio Cruz-López
Unidad Multidisciplinaria de Docencia e Investigación-Sisal, Facultad de Ciencias, Campus Yucatán, Universidad Nacional Autónoma de México (UNAM), Puerto de Abrigo, Sisal, Yucatán, México. pascual.cristina@ciencias.unam.mx. · Unidad Multidisciplinaria de Docencia e Investigación-Sisal, Facultad de Ciencias, Campus Yucatán, Universidad Nacional Autónoma de México (UNAM), Puerto de Abrigo, Sisal, Yucatán, México.
Key points
From the paper's conclusion- The measured blood biochemical values provide a baseline for captive adult common snook and indicate that point-of-care diagnostic devices may serve as practical tools for routine health monitoring of C. undecimalis in the field.
Abstract
The application of clinical biochemistry for health monitoring in marine fish aquaculture remains limited, partly due to the lack of validated analytical protocols and species-specific reference values. This study evaluated the agreement between portable devices (PD) (Accutrend Plus analyzer and protein refractometer) and conventional laboratory methods (clinical diagnostic kits) for quantifying glucose, lactate, cholesterol, triacylglycerols, and total protein in the blood of common snook, Centropomus undecimalis. Thirty-six captive-bred adults (mean body weight: 2.35 ± 0.28 kg and a Fulton index of 0.887 ± 0.036) were sampled. Passing-Bablok regression detected proportional bias for lactate (slope 0.684, 95% CI 0.46-0.92), triacylglycerols (slope 0.594, 95% CI 0.38-0.81), and total protein (slope 0.245, 95% CI 0.13-0.38), with lactate and total protein also showing constant bias; glucose and cholesterol did not deviate from the line of identity. Bland-Altman analysis showed a small mean bias only for lactate (0.024 mg/mL, 6.2%; 95% LoA - 0.010 to 0.058 mg/mL). Glucose showed a moderate though systematic difference (17.2%), whereas cholesterol (- 40.9%), triacylglycerols (35.3%), and particularly total protein (- 55.0%; 95% LoA - 65.88 to - 1.04 mg/mL) showed large systematic differences with LoA that excluded zero. The greatest discrepancy was observed between protein measurement techniques; the refractometer has a lower detection limit that differs considerably from laboratory assessment; therefore, it is not recommended for determining physiological condition. The analysis indicates that portable devices and laboratory techniques yield different absolute results and cannot be used interchangeably. Nevertheless, portable devices offer potential utility for establishing farm-specific baseline levels and identifying elevated lactate or glucose levels associated with stress. The measured blood biochemical values provide a baseline for captive adult common snook and indicate that point-of-care diagnostic devices may serve as practical tools for routine health monitoring of C. undecimalis in the field.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of a portable glucometer for blood glucose measurement in juvenile green sea turtles (Chelonia mydas).Australian veterinary journal · 2026
- Point-of-care blood analyte data in Florida manatees Trichechus manatus latirostris affected by cold stress syndrome and during handling for health assessment procedures.Diseases of aquatic organisms · 2026
- Assessing the reliability and validity for metabolic syndrome diagnosis using point-of-care analyzers in a mobile setup - a pilot study.BMC primary care · 2026
- Análise comparativa de dois reagentes diferentes de glicose de marcas Mindray e Biosystem, mensurados no analisador bioquímico Mindray BS – 120Revista fisio&terapia. · 2026
- Agreement of a Handheld Glucose Meter with Laboratory Serum Glucose Measurements in Dairy CowsDairy · 2026
- Comparative study of blood biochemistry in clinically healthy farmed barramundi ( Lates calcarifer ) and rainbow trout ( Oncorhynchus mykiss ) from freshwater recirculating aquaculture systemsAustralian Veterinary Journal · 2026