Can isolated negative-pressure devices reduce transmission of respiratory infection in critical care units: A laboratory-based controlled experimental observational study.
Sunit Kumar Gupta, Nidhima Aggarwal, Nisha M P, Vamshidhar Chamala, Samarjit Dey, Eeshwar M V, Sumit Bansal, Mukesh Tripathi,
Department of Anesthesiology, AIIMS Mangalagiri, Mangalagiri, India. · Department of Microbiology, AIIMS Mangalagiri, Mangalagiri, India.
Key points
From the paper's conclusion- The negative-pressure isolation device significantly reduced airborne dispersion of bacterial and fungal aerosols compared with plastic sheet canopy barrier methods. These findings support the use of portable negative-pressure systems to enhance aerosol containment during AGPs. Further clinical validation is recommended.
Abstract
BACKGROUND: Healthcare workers in critical care settings are at increased risk of acquiring respiratory infections, particularly during aerosol-generating procedures (AGPs). Portable negative-pressure isolation devices have been proposed to fix airborne infection in isolation rooms. OBJECTIVE: To evaluate whether an isolated negative-pressure device reduces bacterial and fungal aerosol dispersion under controlled laboratory conditions. METHODS: This laboratory-based controlled experimental observational study assessed aerosol containment across three techniques: (1) no cover, (2) plastic sheet canopy barrier, and (3) a portable negative-pressure isolation system. Standardized 5 mL suspensions of Staphylococcus epidermidis and Aspergillus flavus were nebulized on three separate days for each technique. Nutrient agar plates were placed and colony-forming units (CFU) were manually counted. The primary outcome was the mean CFU per technique per day. One-way ANOVA with post-hoc Welch's t-tests (Bonferroni-corrected) was used for statistical comparison. RESULTS: A total of 36 bacterial and fungal plates were analysed. For bacterial aerosols, mean CFU was highest with no cover (535.6 ± 5.38), lower with plastic sheet canopy barrier (429.5 ± 5.41), and lowest with the negative-pressure system (217.5 ± 17.5). Differences were statistically significant (ANOVA p = 9.8 × 10-8; η2 = 0.995). Fungal CFU demonstrated a similar pattern with significantly lower CFU for Technique 3 compared with both other techniques (p < .001). CONCLUSION: The negative-pressure isolation device significantly reduced airborne dispersion of bacterial and fungal aerosols compared with plastic sheet canopy barrier methods. These findings support the use of portable negative-pressure systems to enhance aerosol containment during AGPs. Further clinical validation is recommended.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Exploratory evaluation of negative-pressure operating modes for size- and position-specific aerosol containment during simulated cardiopulmonary resuscitation in a newly developed respiratory isolation chamber.Clinical and experimental emergency medicine · 2026
- Effect of Carbon-Fiber Negative Air Ionization (NAI) on Air Sampled in a Pediatric Hospital: Culturable Airborne Bacteria, Species Diversity, and Biofilm Formation.Microorganisms · 2026
- Bioaerosols in clinical dental settings: an exploratory characterization and evaluation of control strategies.Journal of dentistry · 2026
- Evaluation of Two Cyclonic Bioaerosol Samplers for Hospital Air Monitoring in a Calibrated Test Chamber Integrating Aerodynamics and Biological EndpointsIndoor Air · 2026
- P128. Bacterial viability in PAP airflow after use and reprocessingSLEEP advances · 2026
- MORECOVERY: A Swab-Based Surface Sampling Protocol Incorporating a Nutrient-Free Resuscitation Step for the Detection of Clinically Relevant Gram-Negative Pathogens in the Viable but Non-Culturable State.Bio-protocol · 2026