Exploratory evaluation of negative-pressure operating modes for size- and position-specific aerosol containment during simulated cardiopulmonary resuscitation in a newly developed respiratory isolation chamber.
J H Suh, J Y Hong, K S Sung
Department of Emergency Medicine, Yonsei University College of Medicine, Seoul, Korea. · SS-ENG Co., Ltd., Bucheon, Korea.
Key points
From the paper's conclusion- During simulated CPR, aerosol containment varied by particle size and operator position. Thresholdbased adaptive pressure modulation maintained low residual submicron counts without continuous maximal suction and may offer a balanced strategy by applying additional suction only when needed. Patient-side safety and airflow effects require further evaluation.
Abstract
OBJECTIVE: To compare negative-pressure operating modes for aerosol containment during simulated cardiopulmonary resuscitation (CPR) in a newly developed double-barrier respiratory isolation chamber, focusing on particle-size- and operator-position-specific particle-count patterns. METHODS: In this mannequin-based benchtop simulation, a standardized nebulized aerosol challenge was tested under four modes: 0 Pa, -10 Pa, nominal "Al Pa" adaptive pressure control, and -50 Pa. Particle counts across six size channels were measured at intubator and chest-compressor positions, with and without CPR. Each chamber-applied condition was assessed in 10 re-initialized runs. Analyses were descriptive and estimation-based, using absolute counts, percentage reductions relative to corresponding positive controls, and direct mode comparisons. RESULTS: During CPR, residual particle counts differed by particle size and operator position. Under 0 Pa at the chest-compressor position, mean counts were 199,724.3, 79,870.1, and 12,580.0 for 0.3, 0.5, and 1.0 μm particles; corresponding Al Pa counts were 17,539.3, 1,826.1, and 222.3. Under Al Pa, mean reductions relative to corresponding positive controls ranged from 92.2% to >99.9% across CPR submicron comparisons. Compared with -50 Pa, Al Pa showed 11.1%-30.2% lower mean residual submicron counts. CONCLUSION: During simulated CPR, aerosol containment varied by particle size and operator position. Thresholdbased adaptive pressure modulation maintained low residual submicron counts without continuous maximal suction and may offer a balanced strategy by applying additional suction only when needed. Patient-side safety and airflow effects require further evaluation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Can isolated negative-pressure devices reduce transmission of respiratory infection in critical care units: A laboratory-based controlled experimental observational study.Journal of infection prevention · 2026
- Evaluation of Two Cyclonic Bioaerosol Samplers for Hospital Air Monitoring in a Calibrated Test Chamber Integrating Aerodynamics and Biological EndpointsIndoor Air · 2026
- Downstream particle counting quantifies exposure reduction by miniaturized electrostatic precipitation compared to fibrous filtration.Research square · 2026
- Design assessment and preliminary performance evaluation of a novel device for non-invasive emergency treatment of acute respiratory failure.Biomedical engineering letters · 2026
- Ambient oxygen enrichment during cardiopulmonary resuscitation: an overestimated risk? - Insights from a manikin study.Scandinavian journal of trauma, resuscitation and emergency medicine · 2026
- Positive end-expiratory pressure versus sham valve/zero end-expiratory pressure in cardiopulmonary resuscitation during manual ventilation to improve neurological outcomes in adult patients suffering an out-of-hospital cardiac arrest - an investigator-initiated, pragmatic, registry-based, multicenter, parallel-group, triple-blind randomized controlled superiority clinical trial in the ARREST registry (REVIVE-PEEP protocol Stage-1 Registered Report)medRxiv · 2026