P128. Bacterial viability in PAP airflow after use and reprocessing
O McGuiness, R Salama, S Mathew, E Lambrinos, H Venville, C Menadue, A Piper, S van Hal,
Key points
From the paper's conclusion- These findings suggest a low measurable microbiological burden in PAP device airflow; however, interpretation is limited by the sensitivity of available air sampling techniques.
Abstract
Introduction Positive airway pressure (PAP) devices generate airflow that may contain viable or non-viable microorganisms after clinical use, with potential implications for cross-infection. This study assessed whether airflow from PAP devices contains detectable microbiology. Methods Observational benchtop study of ten PAP devices from users with and without chronic productive cough. Airflow from PAP devices was sampled under pressurised conditions and assessed using a broth-based capture method with subsequent plate culture, with 16S rRNA gene sequencing performed if no growth was detected. Humidifier chamber fluid was analysed using plate cultures and incubated broth. Swabs assessed the device casing, air inlet, air outlet, mask, and tubing before and after reprocessing. Sensitivity testing was performed using nebulised E. coli and P. aeruginosa to define the detection threshold of the sampling system. Results Ten PAP devices (five from people with culture-positive sputum and five end-of-lifespan devices) were assessed. No viable organisms or molecular signals were detected in airflow samples. Surface swabs prior to cleaning demonstrated common human and environmental flora. In sensitivity testing, only highly concentrated, experimentally nebulised bacteria produced detectable culture or molecular signals, indicating high detection threshold for the sampling system. Conclusion In this pilot study, no detectable viable or non-viable microbiology was identified in pressurised air sampled from PAP devices under the conditions tested. These findings suggest a low measurable microbiological burden in PAP device airflow; however, interpretation is limited by the sensitivity of available air sampling techniques. Further experiments using more fastidious organisms are required to better define potential risk.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of Two Cyclonic Bioaerosol Samplers for Hospital Air Monitoring in a Calibrated Test Chamber Integrating Aerodynamics and Biological EndpointsIndoor Air · 2026
- Airway microbial compartmentalization under mechanical ventilation: a randomized pilot study comparing an endotracheal tube with suction and continuous cuff pressure control (Venner PneuX®) to standard intubationRespiratory Research · 2026
- Can isolated negative-pressure devices reduce transmission of respiratory infection in critical care units: A laboratory-based controlled experimental observational study.Journal of infection prevention · 2026
- Intubation and mechanical ventilation lead to an extensive expansion of VAP-associated bacterial species in respiratory secretions of ventilated pigsbioRxiv · 2026
- ESKAPE-Oriented Microbiological Quality Control of Reprocessed Mechanical Ventilators: Implications for Ventilator-Associated Pneumonia and Economic EvaluationMicrobiology Research · 2026
- Personalized automatic management of tracheal cuff pressure and subglottic secretions drainage to prevent pneumonia in critically ill intubated patients. The MICROINHALO multicenter randomized controlled trialIntensive Care Medicine · 2026