Glycemic Concordance of Simplera Sync ™ and Instinct ™ Sensors in MiniMed ™ 780G System Users: A Randomized Crossover Feasibility Study
Maya Laron-Hirsh, Sapir Barel, Anirban Roy, George Mallas, Benyamin Grossman, Robert A. Vigersky, Ohad Cohen, Amir Tirosh
Sheba Medical Center · California State University, Northridge
Key points
From the paper's conclusion- Conclusion: This randomized controlled crossover trial demonstrates that both Simplera Sync and Instinct sensors meet international consensus targets for glycemic control when used with the MM780G system, and that the sensors can be interchanged without meaningful differences in glycemic control.
Abstract
Introduction: This study evaluated the performance of Simplera Sync™ and Instinct™ sensors when paired with the MiniMed™ 780G (MM780G) system and assessed whether switching between sensors affected clinical outcomes of people with type 1 diabetes (PwT1D). Methods: A single-center, two-arm, randomized crossover study in adult PwT1D started with either the Simplera Sync sensor or Instinct sensor for a 6-week period, followed by crossover to the alternate sensor for an additional 6 weeks. The primary end point was noninferiority in time in range (TIR; 70–180 mg/dL), with a noninferiority margin of 7.5%. Secondary end points included HbA1c and other continuous glucose monitoring-derived metrics. Results: Twenty-five participants were assessed for eligibility, 24 were randomized, and 23 completed the study. TIR was 80.6 (9.8)% during the Simplera Sync sensor period and 79.1 (9.5)% during the Instinct sensor period. The mean difference was 1.5% (95% confidence interval: 0.18–2.79), and the noninferiority criteria were met. HbA1c at the end of the Simplera Sync and Instinct sensor periods was 6.8 (0.7)% and 6.7 (0.7)%, respectively. No serious adverse events related to sensor use were reported. Conclusion: This randomized controlled crossover trial demonstrates that both Simplera Sync and Instinct sensors meet international consensus targets for glycemic control when used with the MM780G system, and that the sensors can be interchanged without meaningful differences in glycemic control. This study supports the concept that the automated insulin delivery algorithm, rather than the sensor, is the primary determinant of clinical outcomes in MM780G users.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Prospective multicenter evaluation of glycemic and patient-reported outcomes following transition to a next-generation continuous glucose monitoring system in users of advanced hybrid closed-loop technologyTherapeutic Advances in Endocrinology and Metabolism · 2026
- MiniMed 780G Advanced Hybrid Closed-Loop System Use During Days Without User-Initiated BolusesDiabetes Care · 2026
- In Vitro Dynamic Characterization of a Continuous Glucose Monitoring System Using an Independently Measured Glucose Profile: Amplitude Transfer, Rate Transfer, and Consecutive-Day Agreement.Sensors (Basel, Switzerland) · 2026
- All-in-one sensor design with overlapping sensor transition reduces automode interruptions and improves patient-reported outcomes in adolescents with type 1 diabetes: real-world evidence from simplera sync integration.Diabetes research and clinical practice · 2026
- Simplified Meal Bolus Strategies with Control-IQ+ Automated Insulin Delivery Are Safe and Effective in Adults with Type 2 DiabetesDiabetes Technology & Therapeutics · 2025
- Discordant real‐world glycaemic outcomes with Omnipod™ 5 and MiniMed™ 780G in adults with type 1 diabetes: Why validated measures matterDiabetes Obesity and Metabolism · 2026