The "push-pause" flushing technique in small-bore neonatal catheters: A fluid dynamics analysis.
Fabio Natale, Francesca Spanu, Chiara Ranucci, Roberta Giudice, Giovanni Boscarino, Gianluca Terrin
Department of Maternal and Child Health, Policlinico Umberto I, Sapienza University of Rome, Rome, Italy.
Key points
From the paper's conclusion- In this idealized fluid-dynamics model, the velocity and pressure requirements for reaching conventional Reynolds-number thresholds are highly dependent upon the geometry of the catheter under investigation. These effects are particularly evident in 1 Fr and 2 Fr catheters, where the ΔP are substantially higher. Our results suggest that any potential benefit of the "push-pause" flushing technique in small-bore neonatal catheters would need to rely on hydrodynamic mechanisms other than fully developed turbulence. Further experimental studies of its transient hydrodynamic effects are needed to define safe device-tailored flushing practices.
Abstract
BACKGROUND: Catheter flushing efficiency depends heavily on flow dynamics, as wall shear stress drives deposit removal. The "push-pause" (pulsatile) flushing technique is recommended for the maintenance of vascular access devices (VADs), based on its proposed ability to improve intraluminal clearance through the generation of turbulent flow. However, the hydrodynamic mechanisms underlying this technique remain incompletely characterized, particularly in small-bore neonatal catheters. This study evaluated the theoretical velocity and pressure requirements to generate a steady turbulent flow within neonatal catheters using an idealized fluid-dynamics model. METHODS: Mathematical modeling was applied to polyurethane (PUR) catheters typically used in neonatology, with outer diameters ranging from 1 to 3.5 Fr, internal diameters from 0.17 to 0.80 mm, and lengths from 15 to 40 cm. The Reynolds number (Re) was used to determine the critical velocity (vcrit) corresponding to the onset of transitional flow (Re = 2300) and fully turbulent flow (Re = 4000). The corresponding pressure drops (ΔP) were estimated using the Darcy-Weisbach equation and the Blasius friction factor (λ), assuming 0.9% normal saline at 25°C. RESULTS: At Re = 2300, calculated critical velocities ranged from 2.59 to 12.17 m/s, with ΔP ranging from 0.77 to 59.94 bar. At Re = 4000, calculated velocities ranged from 4.50 to 21.17 m/s, with ΔP ranging from 2.02 to 157.88 bar. The pressure drop in the catheter is directly proportional to its length but inversely proportional to the cube of its internal diameter. CONCLUSIONS: In this idealized fluid-dynamics model, the velocity and pressure requirements for reaching conventional Reynolds-number thresholds are highly dependent upon the geometry of the catheter under investigation. These effects are particularly evident in 1 Fr and 2 Fr catheters, where the ΔP are substantially higher. Our results suggest that any potential benefit of the "push-pause" flushing technique in small-bore neonatal catheters would need to rely on hydrodynamic mechanisms other than fully developed turbulence. Further experimental studies of its transient hydrodynamic effects are needed to define safe device-tailored flushing practices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Irrigation performance of a 2-mm rigid cystoscope designed for neonatal cystoscopy: a bench study.Pediatric surgery international · 2026
- Computational fluid dynamics assessment of a triple-lumen hemodialysis catheter.The journal of vascular access · 2026
- From assumption to assurance: Real-time flow monitoring in neonatal infusions.The journal of vascular access · 2026
- Hydraulic performance of 14-G hemodialysis catheters versus an 8-Fr arterial cannula in a mock 3-kg small-dog cardiopulmonary bypass circuit.PloS one · 2026
- Influencing Factors of Air Bubble-Induced Pressure Difference in Drainage Tubes: An In Vitro Study.Annali italiani di chirurgia · 2026
- Effect of arteriovenous pulsed pneumatic therapy on the hemodynamic microenvironment of autologous arteriovenous fistula in hemodialysis patientsEuropean Journal of Medical Research · 2026