Irrigation performance of a 2-mm rigid cystoscope designed for neonatal cystoscopy: a bench study.
Akio Kawami, Takahiro Jimbo, Tomohiro Aoyama, Kouji Masumoto
Department of Pediatric Surgery, University of Tsukuba Hospital, Tsukuba, Japan. · Department of Pediatric Surgery, University of Tsukuba Hospital, Tsukuba, Japan. jimtaka@md.tsukuba.ac.jp.
Key points
From the paper's conclusion- The 2-mm rigid cystoscope achieved flows up to 40.4 mL/min, within the benchmark range adopted for this benchtop evaluation. These findings support benchtop technical feasibility of the irrigation design. Clinical safety and recommended irrigation pressures remain to be established in closed-system and in vivo studies.
Abstract
PURPOSE: Neonatal and infant cystoscopy is limited by narrow urethral caliber and the lack of sufficiently small cystoscopes with adequate irrigation. We developed a 2-mm (6 Fr) rigid cystoscope prototype and evaluated irrigation performance in a benchtop open-flow system. METHODS: The prototype incorporated a 0.9-mm optical channel, a 0.7-mm working channel, and two irrigation lumens (0.48 mm each). Irrigation performance was assessed under gravity and cuff-assisted irrigation, including temperature-controlled experiments. The maximum cuff pressure of 300 mmHg was used as an upper-bound bench condition and was not intended to represent a recommended clinical setting. Outflow was measured gravimetrically, and mean flow was calculated over 60 s. RESULTS: Gravity irrigation yielded 2.12-8.50 mL/min. Cuff-assisted irrigation increased flow from 17.39 mL/min at 150 mmHg to 40.40 mL/min at 300 mmHg (p < 0.001). At 300 mmHg, warming saline from 20 °C to 37 °C increased flow from 27.49 to 40.40 mL/min (p < 0.001). Probe insertion reduced flow by 4.4% (p = 0.157). CONCLUSION: The 2-mm rigid cystoscope achieved flows up to 40.4 mL/min, within the benchmark range adopted for this benchtop evaluation. These findings support benchtop technical feasibility of the irrigation design. Clinical safety and recommended irrigation pressures remain to be established in closed-system and in vivo studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pressure-flow behavior during RIRS irrigation: a controlled ex vivo porcine kidney study of flow-dependent apparent outflow resistance.World journal of urology · 2026
- The "push-pause" flushing technique in small-bore neonatal catheters: A fluid dynamics analysis.The journal of vascular access · 2026
- A study on the application of operating room nurse-led irrigation-fluid safety management in transurethral ureteropyelolithotomy surgery.Medicine · 2026
- A high-resolution wireless sensing platform for the characterization of urodynamic events in murine models.Biomedical physics & engineering express · 2026
- Wireless MEMS pressure sensing for intrarenal pressure monitoring during ureteroscopy: an in vitro bench-feasibility study.Frontiers in bioengineering and biotechnology · 2026
- UROREC: A cystometry and electromyography recording system for analyzing urinary incontinence parameters in rat models.PloS one · 2026