From assumption to assurance: Real-time flow monitoring in neonatal infusions.
Matheus F P T van Rens, Giovanni Barone, Robin van der Lee, Fiammetta Piersigilli
Neonatal Intensive Care Unit, Radboudumc Amalia Children's Hospital, Nijmegen, The Netherlands. · Neonatal Intensive Care Unit, Infermi Hospital, Rimini, Italy.
Key points
From the paper's conclusion- Improving visibility of patient-end drug delivery may represent future directions for neonatal infusion safety and warrants clinical evaluation.
Abstract
BACKGROUND: In neonatal intensive care, infusion pumps display programmed flow rather than directly confirming drug delivery at the patient-end. At the low infusion rates used in neonatal practice, studies have shown that factors such as compliance, hydrostatic pressure, start-up delay, and partial occlusion can influence actual drug delivery. Although these phenomena are well described, continuous bedside confirmation of patient-end flow isn't available in routine practice. PURPOSE: This editorial discusses the clinical implications of this hidden gap and explores the potential role of real-time flow monitoring. MAIN MESSAGE: We review the mechanisms contributing to flow variability, discuss why conventional pumps cannot verify patient-end drug delivery, and propose a phased pathway for evaluating emerging flow-monitoring technologies. We emphasize that current evidence is based on engineering and early translational studies and that clinical benefit remains to be demonstrated. CONCLUSION: Improving visibility of patient-end drug delivery may represent future directions for neonatal infusion safety and warrants clinical evaluation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Nurses’ perceptions of infusion safety during low-flow neonatal intravenous therapy: An international exploratory surveyJournal of Neonatal Nursing · 2026
- Respiratory function monitoring during neonatal resuscitation: a crossover study of high-fidelity simulated bronchopulmonary dysplasia.Frontiers in pediatrics · 2026
- Anticoagulation management during low flow states in pediatric bivalirudin extracorporeal membrane oxygenation.Frontiers in pediatrics · 2026
- The Accuracy and Precision of Gravity-Driven Infusions.Prehospital emergency care · 2026
- Automated Control of Oxygen Therapy: A State-of-the-Art Review of Closed-Loop Devices, with a Rule-Based, Network-Independent System (EMILY) as a Working Exemplar.Seminars in respiratory and critical care medicine · 2026
- From manual to digital: prospective ICU evaluation of the FIZE kUO® for continuous urine output measurement.BMC nursing · 2026