Biomechanical effects of far-cortex over-penetration in locking plate fixation: A finite element study.
Natig Valiyev, Cihan Zaman
Department of Orthopaedics and Traumatology, Ümraniye Training and Research Hospital, Istanbul, Turkey. · Department of Orthopaedics and Traumatology, Sarıkamış State Hospital, Kars, Turkey.
Key points
From the paper's conclusion- This study quantified the biomechanical consequences of incremental far-cortex over-penetration at terminal locking screw positions in a locking compression plate construct using a validated three-dimensional finite element model of a femoral diaphyseal fracture (AO/OTA 32-A3).
Abstract
This study quantified the biomechanical consequences of incremental far-cortex over-penetration at terminal locking screw positions in a locking compression plate construct using a validated three-dimensional finite element model of a femoral diaphyseal fracture (AO/OTA 32-A3). Four terminal screw engagement configurations were evaluated: unicortical fixation, fully contained bicortical fixation, 1-thread over-penetration (0.6 mm), and 3-thread over-penetration (1.8 mm). Axial compression (1000 N), torsional loading (±10 Nm), and cyclic dynamic loading (100-1000 N, 1 × 106 cycles) simulating the 8-week early postoperative period were applied. Model validation against published experimental data demonstrated agreement within 4% for all key biomechanical parameters. Fully contained bicortical fixation was the only configuration to remain below the predicted fatigue threshold, demonstrating the lowest peak von Mises stress (74.8 MPa) and a Miner's cumulative damage index of D = 0.291. Unicortical fixation (D = 1.103), 1-thread over-penetration (D = 1.362), and 3-thread over-penetration (D = 2.968) all exceeded the failure threshold (D ≥ 1.0). Under the loading conditions and material assumptions of the present finite element model, unicortical terminal fixation alone did not sustain simulated cyclic physiological loading during the pre-callus phase without exceeding the predicted fatigue damage threshold. These findings provide preliminary, model-dependent computational evidence suggesting that fully contained bicortical fixation may offer greater fatigue resistance, and that the biomechanical basis of current terminal screw placement guidelines warrants further experimental investigation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Investigation of optimal placement of locking attachment plate in periprosthetic femoral fractures : a biomechanical comparative study.Bone & joint research · 2026
- Biomechanical and finite element analysis of a novel annular multi-axial locking plate for patellar fracture fixation.Journal of orthopaedic surgery and research · 2026
- Biomechanical evaluation of a fully cortical-threaded screw for modified cortical bone trajectory fixation: Combined experimental and finite element studyProceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine · 2026
- Optimizing Volar Lunate Facet Fracture Fixation: Finite Element Analysis and Cadaveric Testing.Journal of orthopaedic research : official publication of the Orthopaedic Research Society · 2026
- Biomechanical Study of the Preventive Effect of Different Cephalomedullary Fixation Methods on the Occurrence of Femoral Neck Fractures in Osteoporotic Femurs.Medicina (Kaunas, Lithuania) · 2026
- Cage-Orientation-Dependent Biomechanics of Endoscopic-Compatible Bidirectional Cortical Bone Trajectory-Modified Translaminar Facet Screw Fixation versus Bilateral Percutaneous Pedicle Screw Fixation in Osteoporotic TLIF: A Finite Element Study.Medical engineering & physics · 2026